5 Block USER INTERFACE
36
5.1.3
Function group OPERATION
→
USER INTERFACE
C
TEST DISPLAY
2040
MODBUS register:
2513
Data type:
Integer
Access:
read/write
Device Functions Proline Promass 83 MODBUS RS485
CONTROL
Function description
USER INTERFACE → CONTROL → OPERATION
For testing the operability of the local display and its pixels.
Options:
0 = OFF
1 = ON
Factory setting: OFF
Test sequence:
1. Start the test by selecting "ON".
2. All pixels of the main line, additional line and information line are darkened
for minimum 0.75 seconds.
3. Main line, additional line and information line show an "8" in each field for
minimum 0.75 seconds.
4. Main line, additional line and information line show a "0" in each field for
minimum 0.75 seconds.
5. Main line, additional line and information line show nothing (blank display)
for minimum 0.75 second.
When the test is completed, the local display returns to its initial state and the
setting changes to "OFF".
→
CAA
BASIC CONFIGURATION
UNLOCKING/LOCKING
OPERATION
Endress+Hauser
200
202
204