Commissioning Instructions for PTB gas approval
DE-08-MI002-PTB007
Proline Promass 84
Coriolis gas meter for applications subject to legal metrological
control
SD00128D/06/EN/13.11
71144260
Valid as of version
HART: V 3.01.XX (device software)
Modbus: V 3.06.XX (device software)

Advertisement

Table of Contents
loading

Summary of Contents for Endress+Hauser Proline Promass 84

  • Page 1 Commissioning Instructions for PTB gas approval DE-08-MI002-PTB007 Proline Promass 84 Coriolis gas meter for applications subject to legal metrological control SD00128D/06/EN/13.11 71144260 Valid as of version HART: V 3.01.XX (device software) Modbus: V 3.06.XX (device software)
  • Page 3: Table Of Contents

    Commissioning Instructions for PTB gas approval Proline Promass 84 Table of contents Table of contents 7.13 User interface Information line Multiplex . . . 31 General information ....4 7.14 Totalizer Totalizer 1 to 3 Configuration .
  • Page 4: General Information

    • for the Proline Promass 84 HART measuring device: – BA00109D Operating Instructions – BA00110D Description of Device Functions • for the Proline Promass 84 Modbus RS485 measuring device: – BA00129D Operating Instructions – BA00130D Description of Device Functions Components of the measuring system 1.2.1...
  • Page 5: Identification

    Commissioning Instructions for PTB gas approval Proline Promass 84 Identification Identification The following options are available for identification of the measuring device:: • Nameplate specifications • Order code with breakdown of the device features on the delivery note • Enter serial numbers from nameplates in W@M Device Viewer (www.endress.com/deviceviewer): All information about the measuring device is displayed.
  • Page 6: Device Designation And Versions

    Identification Commissioning Instructions for PTB gas approval Proline Promass 84 Device designation and versions The "Promass 84" Coriolis gas meter consists of the following parts: • Transmitter: Promass 84 • Sensor: Promass A, Promass F, Promass X and Promass O The "Promass 84"...
  • Page 7: Nameplates

    Commissioning Instructions for PTB gas approval Proline Promass 84 Identification Nameplates The nameplates are attached to the transmitter and sensor and describe the most important technical information pertaining to the measuring device in question. The following sections describe the structure of the various nameplates using examples: 2.2.1...
  • Page 8 Identification Commissioning Instructions for PTB gas approval Proline Promass 84 2.2.2 Additional nameplate for gas meters subject to legal metrological control Environmental Class Accuracy Class Qmin = Qmax = pe, min = pe, max = Gas Temperature Ambient Temperature Gas Type Pulse Value 1259 DE–08–MI002–PTB007...
  • Page 9 Commissioning Instructions for PTB gas approval Proline Promass 84 Identification 2.2.3 Sensor nameplate (remote version) Order Code: Ser.No.: Ext. ord. cd.: K-factor: Materials: Density cal: A0011002 Fig. 3: Example of a sensor nameplate Certificate recipient Name of the transmitter Order code, meaning of individual letters and digits ä 11 Serial number (Ser.
  • Page 10 Identification Commissioning Instructions for PTB gas approval Proline Promass 84 2.2.4 Nameplate for connections active See operating manual passive Betriebsanleitung beachten normally open contact Observer manuel d'instruction normally closed contact Ser.No.: L1/L+ Supply / Versorgung / N/L- Tension d'alimentation Update 1 Update 2 ex works / ab Werk / réglages usine...
  • Page 11 Commissioning Instructions for PTB gas approval Proline Promass 84 Identification 2.2.5 Structure of order code/serial number The order code/serial number (type code) describes the exact structure and equipment level of the measuring system. It can be found on the nameplate of the transmitter and sensor and is structured as follows: 8 4 * * * –...
  • Page 12: Wiring

    Wiring Warning! When connecting Ex-certified devices, see the notes and diagrams in the Ex-specific supplement to these operating instructions. Please do not hesitate to contact your Endress+Hauser sales office if you have any questions. Terminal assignment The values for the individual inputs and outputs can be found in the Technical Data ä 22 Terminal no.
  • Page 13: Electrical Connection Of Measuring Unit

    Commissioning Instructions for PTB gas approval Proline Promass 84 Wiring Electrical connection of measuring unit HART –/b –/b Modbus RS485 a/a b/b –/c HART Modbus RS485 – 27 A (RxD/TxD-N) + 26 B (RxD/TxD-P) – 25 – + 24 –...
  • Page 14: Electrical Connection Of Remote Version

    Wiring Commissioning Instructions for PTB gas approval Proline Promass 84 Electrical connection of remote version S1 S1 S2 S2 GND TM TM TT TT 10 11 12 41 42 10 11 12 41 42 S1 S1 S2 S2 GND TM TM TT TT a0003681 Fig.
  • Page 15: Operation And Commissioning

    • HART protocol or Modbus RS485 communication (depending on measuring device version) Exact information on these topics can be found in the accompanying operating instructions (BA00109D, Proline Promass 84 HART or BA00129D, Proline Promass 84 Modbus RS485). Hardware configurations Using hardware switches, different configurations can be set up depending on the measuring device version: •...
  • Page 16: Operation Subject To Legal Metrological Control ("Custody Transfer")

    All Promass 84 Coriolis gas meters are tested at Endress+Hauser Flowtec AG or on an approved reference test rig (e.g. PIGSAR) using reference measurements. The devices are placed on the market and put into use by either Endress+Hauser Flowtec AG in accordance with Measuring Instruments Directive, Module D or by an notified body in accordance with Module F.
  • Page 17: Requirements For Usage

    Commissioning Instructions for PTB gas approval Proline Promass 84 Operation subject to legal metrological control ("Custody Requirements for usage The meter must be operated with an uninterruptible voltage supply (emergency current supply in according to EN 60654-2), which can bridge a power failure and secure operation for at least 3 days or until maintenance operations are is carried out.
  • Page 18: Zero Point Adjustment

    – When the zero point adjustment completes, the "ZERO ADJUST." function reappears on the display. Back to the HOME position: – Press and hold down the Esc keys ( X ) for longer than three seconds, or repeatedly press and release these keys. Endress+Hauser...
  • Page 19: Configuring The Measuring Device For Operation Subject To Legal Metrological Control

    Commissioning Instructions for PTB gas approval Proline Promass 84 Operation subject to legal metrological control ("Custody Configuring the measuring device for operation subject to legal metrological control Precondition: The device is operational and not in custody transfer mode. Configuration of those functions which are important for custody transfer measurement, such as output configuration, custody transfer variable and measuring mode.
  • Page 20: Sealing Of The Measuring Device

    Operation subject to legal metrological control ("Custody transfer")Commissioning Instructions for PTB gas approval Proline Promass Sealing of the measuring device a0001778 Fig. 7: Examples of how to seal the various device versions. Endress+Hauser...
  • Page 21: Unlocking The Measuring Device

    Commissioning Instructions for PTB gas approval Proline Promass 84 Operation subject to legal metrological control ("Custody Unlocking the measuring device Precondition: The device is operational and is already set to custody transfer mode. Disconnect the device from the operating voltage.
  • Page 22: Technical Data

    Operating programs FieldCare FieldCare is Endress+Hauser’s FDT-based plant asset management tool and allows the configuration and diagnosis of intelligent field devices. The use of status information means that you also have a simple but effective tool for monitoring devices. Proline flowmeters are accessed via a service interface or the FXA195 or FXA 193.
  • Page 23 Commissioning Instructions for PTB gas approval Proline Promass 84 Technical data 6.1.5 Input Measured variable • Mass flow (proportional to the phase difference between two sensors mounted on the measuring tube which register a phase shift in the oscillation) • Fluid density (proportional to the resonance frequency of the measuring tube) •...
  • Page 24 Technical data Commissioning Instructions for PTB gas approval Proline Promass 84 Input signal Status input (auxiliary input), Promass 84 HART U = 3 to 30 V DC, R = 5 k, galvanically isolated. Configurable for: totalizer reset, positive zero return, error message reset, start zero point...
  • Page 25 Commissioning Instructions for PTB gas approval Proline Promass 84 Technical data Modbus RS485 • Modbus device type: slave • Address range: 1 to 247 • Supported function codes: 03, 04, 06, 08, 16, 23 • Broadcast: supported by the function codes 06, 16, 23 •...
  • Page 26 Technical data Commissioning Instructions for PTB gas approval Proline Promass 84 6.1.8 Normal operating conditions • Measured variable: mass of gas flowing through in kg, volume of gas flowing through in m (possible with clean gases) • Gas types, covered by MI-002 : fuel gases, technically pure flammable gases and mixtures of these gases in a gaseous state •...
  • Page 27 CE mark The measuring system is in conformity with the statutory requirements of the EC directives. Endress+Hauser confirms successful testing of the device by affixing to it the CE mark. C-tick mark The measuring system meets the EMC requirements of the Australian Communications and Media Authority (ACMA).
  • Page 28: Functions Relevant To Custody Transfer

    Options: YES CT signal. (if relevant to custody transfer) * available only for Proline Promass 84 HART measuring device ** available only for Proline Promass 84 Modbus RS485 measuring device Meas. variables System units Configuration Function name Description Configuration UNIT 0400 Use this function to select the desired unit.
  • Page 29: Meas. Variables System Units

    1004 Start setup. – GAS MEASUREMENT QUICK SETUP 1006 Start setup. – COMMUNICATION ** T-DAT 1006 Saving from EEPROM to T-DAT. – SAVE/LOAD Loading from T-DAT to EEPROM. ** available only for the Proline Promass 84 Modbus RS485 measuring device Endress+Hauser...
  • Page 30: Display Control Basic Config

    Functions relevant to custody transfer Commissioning Instructions for PTB gas approval Proline Promass 84 Display Control Basic config. Function name Description Configuration DISPLAY DAMPING 2002 Time constant which determines how the display reacts User input: to severely fluctuating flow variables.
  • Page 31: User Interface Additional Line Multiplex

    Commissioning Instructions for PTB gas approval Proline Promass 84 Functions relevant to custody transfer User interface Additional line Multiplex 7.11 Function name Description Configuration ASSIGN 2420 Use this function to assign a display value to the main Options: line. e.g. MASS FLOW...
  • Page 32: Totalizer Handling Totalizer

    Functions relevant to custody transfer Commissioning Instructions for PTB gas approval Proline Promass 84 Totalizer Handling totalizer 7.15 Function name Description Configuration RESET ALL TOTALIZERS 3800 Use this function to reset the total and the overflow of Options: all totalizers to zero.
  • Page 33: Outputs Pulse/Freq. Output Configuration

    Commissioning Instructions for PTB gas approval Proline Promass 84 Functions relevant to custody transfer Outputs Pulse/freq. output Configuration 7.18 Function name Description Configuration MODE OF OPERATION 4200 Configuration of output as: Options: Frequency output, ä 33 e.g. PULSE Pulse output, ä 33 Status output, ä...
  • Page 34 Functions relevant to custody transfer Commissioning Instructions for PTB gas approval Proline Promass 84 7.18.3 Status output Function name Description Configuration ASSIGN STATUS 4241 Use this function to assign a switching function to the Options: status output. e.g. FAULT MESSAGE...
  • Page 35: Outputs Pulse/Freq. Output Operation

    Commissioning Instructions for PTB gas approval Proline Promass 84 Functions relevant to custody transfer Outputs Pulse/freq. output Operation 7.19 7.19.1 Frequency output Function name Description Configuration ACTUAL FREQUENCY 4301 Use this function to display the frequency currently Display: output. e.g. 0 Hz...
  • Page 36: Outputs Relay Output Configuration

    Functions relevant to custody transfer Commissioning Instructions for PTB gas approval Proline Promass 84 Outputs Relay output Configuration 7.20 Function name Description Configuration ASSIGN RELAY 4700 Use this function to assign a switching function to the Options: relay output. e.g. FAULT MESSAGE...
  • Page 37: Basic Function Process Parameter Configuration

    Commissioning Instructions for PTB gas approval Proline Promass 84 Functions relevant to custody transfer Basic function Process parameter Configuration 7.24 Function name Description Configuration ASSIGN LOW FLOW CUT 6400 Use this function to assign a switching function to the Options: low flow cut off.
  • Page 38: Basic Function Process Parameter Adjustment

    Functions relevant to custody transfer Commissioning Instructions for PTB gas approval Proline Promass 84 Basic function Process parameter Adjustment 7.27 Function name Description Configuration ZERO POINT ADJUST 6480 Use this function to activate zero point adjustment. Options: e.g. CANCEL DENSITY ADJUST...
  • Page 39: Basic Function Sensor Data Flow Coeff

    Commissioning Instructions for PTB gas approval Proline Promass 84 Functions relevant to custody transfer Basic function Sensor data Flow coeff. 7.31 Function name Description Configuration TEMPERATURE 6840 Use this function to display the temperature coefficient display only COEFFICIENT KM TEMP. COEFFICIENT KM 2...
  • Page 40: Supervision System Operation

    Functions relevant to custody transfer Commissioning Instructions for PTB gas approval Proline Promass 84 Supervision System Operation 7.34 Function name Description Configuration SIMULATION FAILSAFE 8042 Use this function to switch all inputs, outputs and Options: MODE totalizers to their respective failsafe mode.
  • Page 41: Supervision Version-Info I/O Module

    Commissioning Instructions for PTB gas approval Proline Promass 84 Functions relevant to custody transfer Supervision Version-info I/O module 7.38 Function name Description Configuration I/O MODULE TYPE 8300 Use this function to display the configuration of the I/O display only module complete with terminal number.
  • Page 42: Service & Analysis Application Gas Measurement

    Functions relevant to custody transfer Commissioning Instructions for PTB gas approval Proline Promass 84 Service & Analysis Application Gas measurement 7.41 Function name Description Configuration GAS MODE 9200 Use this function to select the gas mode for gas Options: MODE 1...
  • Page 43 Commissioning Instructions for PTB gas approval Proline Promass 84 Functions relevant to custody transfer Endress+Hauser...
  • Page 44 www.endress.com/worldwide SD00128D/06/EN/13.11 71144260 FM+SGML9.0 ProMoDo...