2.1.5 Codes identifying cells ....7 7.4.2 Function group FLOW COEFFICIENT ..62 Function matrix Proline Promass 83 ... . 8 7.4.3 Function group DENSITY COEFFICIENT . . 63 7.4.4...
Page 4
Table of contents for local operation Device Functions Proline Promass 83 FOUNDATION Fieldbus Factory settings ..... . . 10.1 SI units (not for USA and Canada) ... . . 90 10.1.1 Low flow cut off, full scale value,...
Device Functions Proline Promass 83 FOUNDATION Fieldbus 1 Notes on using this manual Notes on using this manual There are various ways of locating the description of a function of your choice in the manual: Using the table of contents to locate a function description The designations of all the cells in the function matrix are listed in the table of contents.
2 Function matrix Device Functions Proline Promass 83 FOUNDATION Fieldbus Function matrix General layout of the function matrix The function matrix consists of four levels: Blocks → Groups → Function groups → Functions 0000 0001 0002 0003 0009 … 0400...
Device Functions Proline Promass 83 FOUNDATION Fieldbus 2 Function matrix 2.1.5 Codes identifying cells Each cell (block, group, function group and function) in the function matrix has an individual, unique code. Blocks: The code is a letter (A, B, C, etc.) Groups: The code consists of three letters (AAA, ABA, BAA, etc.).
2 Function matrix Device Functions Proline Promass 83 FOUNDATION Fieldbus Function matrix Proline Promass 83 FUNCTION BLOCKS GROUPS GROUPS → → MEASURED VARIABLES MEASURING VALUES page 10 → ( → Page 9) SYSTEM UNITS page 14 ↓ → SPECIAL UNITS page 14 →...
3 Block MEASURED VARIABLES Device Functions Proline Promass 83 FOUNDATION Fieldbus Group MEASURING VALUES 3.1.1 Function group MAIN VALUES MEASURED ⇒ ⇒ MEASURING VALUES MAIN VALUES VARIABLES Function description MEASURED VARIABLES → MEASURING VALUES → MAIN VALUES Note! • The engineering units of all the measured variables shown here can be set in the SYSTEM UNITS group.
3 Block MEASURED VARIABLES Device Functions Proline Promass 83 FOUNDATION Fieldbus Function description MEASURED VARIABLES → MEASURING VALUES → ADDITIONAL VALUES CORRECTED Note! TARGET VOLUME FLOW This function is not available unless % MASS / % VOLUME was selected in the (0024) DENSITY FUNCTION function (7000).
Device Functions Proline Promass 83 FOUNDATION Fieldbus 3 Block MEASURED VARIABLES Function description MEASURED VARIABLES → MEASURING VALUES → ADDITIONAL VALUES CORRECTED CARRIER Note! VOLUME FLOW This function is not available unless % MASS / % VOLUME was selected in the (0029) DENSITY FUNCTION function (7000).
3 Block MEASURED VARIABLES Device Functions Proline Promass 83 FOUNDATION Fieldbus Group SYSTEM UNITS 3.2.1 Function group CONFIGURATION MEASURED ⇒ MEASURING VALUES VARIABLES ⇓ ⇒ SYSTEM UNITS CONFIGURATION Function description MEASURED VARIABLES → SYSTEM UNITS → CONFIGURATION You can select the units for measured variables in this function group.
Page 15
Device Functions Proline Promass 83 FOUNDATION Fieldbus 3 Block MEASURED VARIABLES Function description MEASURED VARIABLES → SYSTEM UNITS → CONFIGURATION UNIT VOLUME FLOW Use this function to select the unit for displaying the volume flow (volume/time). (0402) The unit you select here is also valid for: •...
Page 16
3 Block MEASURED VARIABLES Device Functions Proline Promass 83 FOUNDATION Fieldbus Function description MEASURED VARIABLES → SYSTEM UNITS → CONFIGURATION UNIT CORRECTED Use this function to select the unit for displaying the corrected volume flow VOLUME FLOW (corrected volume/time). (0404) The unit you select here is also valid for: •...
Device Functions Proline Promass 83 FOUNDATION Fieldbus 3 Block MEASURED VARIABLES 3.2.2 Function group ADDITIONAL CONFIGURATION MEASURED ⇒ MEASURING VALUES VARIABLES ⇓ ⇒ SYSTEM UNITS CONFIGURATION ⇓ ADD. CONFIGURATION 042 Function description MEASURED VARIABLES → SYSTEM UNITS → ADDITIONAL CONFIGURATION Note! The units selected here have no effect on the fieldbus.
3 Block MEASURED VARIABLES Device Functions Proline Promass 83 FOUNDATION Fieldbus Function description MEASURED VARIABLES → SYSTEM UNITS → ADDITIONAL CONFIGURATION UNIT LENGTH Use this function to select the unit for displaying the length of the nominal diameter. (0424) The unit you select here is valid for: •...
Factory setting: Note! You will find a flowchart of the GAS MEASUREMENT setup menu on page 21. For more information on Setup menus, please refer to the Operating Instructions Proline Promass 83, BA065D/06/en. T-DAT SAVE/LOAD Use this function to save the parameter settings / configuration of the transmitter in a...
Device Functions Proline Promass 83 FOUNDATION Fieldbus 4 Block QUICK SETUP Quick Setup "Gas Measurement" Note! The procedure is described in detail in the Operating Instructions for Proline Promass 83 FOUNDATION Fieldbus, BA065D/06/en. XXX.XXX.XX HOME POSITION 1004 Quick Setup Setup...
• If you press the O/S keys at startup, the language defaults to "ENGLISH". • You can change the language group via the configuration software ToF Tool- Fieldtool Package. Please do not hesitate to contact your Endress+Hauser sales office if you have any questions.
• You can also disable programming in this function by entering any number (other than the defined private code). • The Endress+Hauser service organization can be of assistance if you mislay your personal code. • In this function, access to programming is only valid for local configuration. If functions or parameters are to be changed via the fieldbus, programming must be enabled separately in the parameter "Access - code"...
Device Functions Proline Promass 83 FOUNDATION Fieldbus 5 Block USER INTERFACE Function description USER INTERFACE → CONTROL → UNLOCKING/LOCKING DEFINE Use this function to specify a personal code for enabling programming in the function PRIVATE CODE ACCESS CODE. (2021) User input: 0 to 9999 (max.
5 Block USER INTERFACE Device Functions Proline Promass 83 FOUNDATION Fieldbus 5.1.3 Function group OPERATION ⇒ ⇒ USER INTERFACE CONTROL BASIC CONFIGURATION 200 ⇓ UNLOCKING/LOCKING 202 ⇓ OPERATION Function description USER INTERFACE → CONTROL → OPERATION TEST DISPLAY Use this function to test the operability of the local display and its pixels.
Device Functions Proline Promass 83 FOUNDATION Fieldbus 5 Block USER INTERFACE Group MAIN LINE 5.2.1 Function group CONFIGURATION ⇒ USER INTERFACE CONTROL ⇓ ⇒ MAIN LINE CONFIGURATION Function description USER INTERFACE → MAIN LINE → CONFIGURATION A0001253 1 = main line, 2 = additional line, 3 = information line...
Page 28
5 Block USER INTERFACE Device Functions Proline Promass 83 FOUNDATION Fieldbus Function description USER INTERFACE → MAIN LINE → CONFIGURATION ASSIGN Advanced options with optional software package CONCENTRATION: (continued) TARGET MASS FLOW % TARGET MASS FLOW TARGET VOLUME FLOW % TARGET VOLUME FLOW...
Device Functions Proline Promass 83 FOUNDATION Fieldbus 5 Block USER INTERFACE 5.2.2 Function group MULTIPLEX ⇒ USER INTERFACE CONTROL ⇓ ⇒ MAIN LINE CONFIGURATION ⇓ MULTIPLEX Function description USER INTERFACE → MAIN LINE → MULTIPLEX ASSIGN Use this function to define the second reading to be displayed in the main line (2220) alternately (every 10 seconds) with the value defined in the ASSIGN function (2200).
Page 30
5 Block USER INTERFACE Device Functions Proline Promass 83 FOUNDATION Fieldbus Function description USER INTERFACE → MAIN LINE → MULTIPLEX ASSIGN Advanced options with optional software package ADVANCED DIAGNOSTICS: (continued) MASS FLOW DEVIATION DENSITY DEVIATION REFERENCE DENSITY DEVIATION TEMPERATURE DEVIATION...
Device Functions Proline Promass 83 FOUNDATION Fieldbus 5 Block USER INTERFACE Group ADDITION LINE 5.3.1 Function group CONFIGURATION ⇒ USER INTERFACE CONTROL ⇓ MAIN LINE ⇓ ⇒ ADDITION LINE CONFIGURATION Function description USER INTERFACE → ADDITION LINE → CONFIGURATION A0001253...
Page 32
5 Block USER INTERFACE Device Functions Proline Promass 83 FOUNDATION Fieldbus Function description USER INTERFACE → ADDITION LINE → CONFIGURATION ASSIGN Advanced options with optional software package CONCENTRATION: (continued) TARGET MASS FLOW % TARGET MASS FLOW TARGET VOLUME FLOW % TARGET VOLUME FLOW...
Page 33
Device Functions Proline Promass 83 FOUNDATION Fieldbus 5 Block USER INTERFACE Function description USER INTERFACE → ADDITION LINE → CONFIGURATION FORMAT Note! (2402) This function is not available unless a number was selected in the ASSIGN function (2400). Use this function to define the maximum number of places after the decimal point displayed for the reading in the additional line.
5 Block USER INTERFACE Device Functions Proline Promass 83 FOUNDATION Fieldbus 5.3.2 Function group MULTIPLEX ⇒ USER INTERFACE CONTROL ⇓ MAIN LINE ⇓ ⇒ ADDITION LINE CONFIGURATION ⇓ MULTIPLEX Function description USER INTERFACE → ADDITION LINE → MULTIPLEX ASSIGN Use this function to define the second reading to be displayed in the additional line (2420) alternately (every 10 seconds) with the value defined in the ASSIGN function (2400).
Page 35
Device Functions Proline Promass 83 FOUNDATION Fieldbus 5 Block USER INTERFACE Function description USER INTERFACE → ADDITION LINE → MULTIPLEX ASSIGN Advanced options with optional software package ADVANCED DIAGNOSTICS: (continued) MASS FLOW DEVIATION DENSITY DEVIATION REFERENCE DENSITY DEVIATION TEMPERATURE DEVIATION...
Page 36
5 Block USER INTERFACE Device Functions Proline Promass 83 FOUNDATION Fieldbus Function description USER INTERFACE → ADDITION LINE → MULTIPLEX DISPLAY MODE Note! (2423) This function is not available unless one of the following was selected in the ASSIGN function (2420): •...
Device Functions Proline Promass 83 FOUNDATION Fieldbus 5 Block USER INTERFACE Group INFORMATION LINE 5.4.1 Function group CONFIGURATION ⇒ USER INTERFACE CONTROL ⇓ MAIN LINE ⇓ ADDITION LINE ⇓ ⇒ INFORMATION LINE CONFIGURATION Function description USER INTERFACE → INFORMATION LINE → CONFIGURATION...
5 Block USER INTERFACE Device Functions Proline Promass 83 FOUNDATION Fieldbus Function description USER INTERFACE → INFORMATION LINE → CONFIGURATION ASSIGN Advanced options with optional software package CONCENTRATION: (continued) TARGET MASS FLOW % TARGET MASS FLOW TARGET VOLUME FLOW % TARGET VOLUME FLOW...
Device Functions Proline Promass 83 FOUNDATION Fieldbus 5 Block USER INTERFACE Function description USER INTERFACE → INFORMATION LINE → CONFIGURATION FORMAT Note! (2602) This function is not available unless a number was selected in the ASSIGN function (2600). Use this function to define the maximum number of places after the decimal point displayed for the reading in the information line.
5 Block USER INTERFACE Device Functions Proline Promass 83 FOUNDATION Fieldbus 5.4.2 Function group MULTIPLEX ⇒ USER INTERFACE CONTROL ⇓ MAIN LINE ⇓ ADDITION LINE ⇓ ⇒ INFORMATION LINE CONFIGURATION ⇓ MULTIPLEX Function description USER INTERFACE → INFORMATION LINE → MULTIPLEX...
Device Functions Proline Promass 83 FOUNDATION Fieldbus 5 Block USER INTERFACE Function description USER INTERFACE → INFORMATION LINE → MULTIPLEX ASSIGN Advanced options with optional software package CONCENTRATION: (2620) TARGET MASS FLOW (Continued) % TARGET MASS FLOW TARGET VOLUME FLOW...
5 Block USER INTERFACE Device Functions Proline Promass 83 FOUNDATION Fieldbus Function description USER INTERFACE → INFORMATION LINE → MULTIPLEX FORMAT Note! (2622) This function is not available unless a number was selected in the ASSIGN function (2620). Use this function to define the maximum number of places after the decimal point for the second value displayed in the information line.
6 Block TOTALIZER Device Functions Proline Promass 83 FOUNDATION Fieldbus Group TOTALIZERS (1 to 3) 6.1.1 Function group CONFIGURATION ⇒ ⇒ TOTALIZER TOTALIZER 1 CONFIGURATION ⇓ ⇒ TOTALIZER 2 CONFIGURATION ⇓ ⇒ TOTALIZER 3 CONFIGURATION Function description TOTALIZER → TOTALIZERS (1 to 3) → CONFIGURATION The function descriptions below apply to totalizers 1 to 3;...
Device Functions Proline Promass 83 FOUNDATION Fieldbus 6 Block TOTALIZER Function description TOTALIZER → TOTALIZERS (1 to 3) → CONFIGURATION UNIT TOTALIZER Use this function to define the unit for the totalizer's measured variable, as selected (3001) beforehand. Options (for the MASS FLOW assignment): Metric →...
Factory setting: Note! Hardware write protection is activated and deactivated by means of a jumper on the I/O board (see Operating Instructions for Proline Promass 83 FOUNDATION Fieldbus BA 065D/06/en). SIMULATION Use this function to check whether a simulation in the Analog Input function block is (6201) possible.
7 Block BASIC FUNCTION Device Functions Proline Promass 83 FOUNDATION Fieldbus 7.1.2 Function group FUNCTION BLOCKS ⇒ ⇒ BASIC FUNCTION FOUNDATION FIELDBUS GGA CONFIGURATION ⇓ FUNCTION BLOCKS Function description BASIC FUNCTION → FOUNDATION FIELDBUS → FUNCTION BLOCKS BLOCK SELECTION In this function, a function block can be selected, whose value and status is shown in (6220) the following functions.
Device Functions Proline Promass 83 FOUNDATION Fieldbus 7 Block BASIC FUNCTION 7.1.3 Function group INFORMATION ⇒ ⇒ BASIC FUNCTION FOUNDATION FIELDBUS GGA CONFIGURATION ⇓ FUNCTION BLOCKS ⇓ INFORMATION Function description BASIC FUNCTION → FOUNDATION FIELDBUS → INFORMATION MANUFACTURER ID Use this function to view the manufacturer ID in decimal numerical format.
7 Block BASIC FUNCTION Device Functions Proline Promass 83 FOUNDATION Fieldbus Group PROCESS PARAMETER 7.2.1 Function group CONFIGURATION ⇒ BASIC FUNCTION FOUNDATION FIELDBUS GGA ⇓ ⇒ PROCESS PARAMETER CONFIGURATION Function description BASIC FUNCTION → PROCESS PARAMETER → CONFIGURATION ASSIGN LOW FLOW Use this function to assign the switching point for low flow cut off.
Device Functions Proline Promass 83 FOUNDATION Fieldbus 7 Block BASIC FUNCTION Function description BASIC FUNCTION → PROCESS PARAMETER → CONFIGURATION PRESSURE SHOCK The closure of a valve can cause brief but severe movements of the fluid in the piping SUPPRESSION system, movements which the measuring system registers.
7 Block BASIC FUNCTION Device Functions Proline Promass 83 FOUNDATION Fieldbus 7.2.2 Function group EPD PARAMETER ⇒ BASIC FUNCTION FOUNDATION FIELDBUS GGA ⇓ ⇒ PROCESS PARAMETER CONFIGURATION ⇓ EPD PARAMETER Function description BASIC FUNCTION → PROCESS PARAMETER → EPD PARAMETER EMPTY PIPE Use this function to activate the empty pipe detection (EPD).
Device Functions Proline Promass 83 FOUNDATION Fieldbus 7 Block BASIC FUNCTION Function description BASIC FUNCTION → PROCESS PARAMETER → EPD PARAMETER EPD EXCITATION Use this function to activate the empty pipe detection (EPD). CURRENT MAXIMUM In the event of inhomogeneous fluids or air bubbles, the excitation current of the (6426) measuring pipes increases.
7 Block BASIC FUNCTION Device Functions Proline Promass 83 FOUNDATION Fieldbus Function description BASIC FUNCTION → PROCESS PARAMETER → REFERENCE PARAMETER EXPANSION Note! COEFFICIENT This function is not available unless the CALCULATED REFERENCE DENSITY setting (6462) was selected in the CORRECTED VOLUME CALCULATION function (6460).
Device Functions Proline Promass 83 FOUNDATION Fieldbus 7 Block BASIC FUNCTION 7.2.4 Function group ADJUSTMENT ⇒ BASIC FUNCTION FOUNDATION FIELDBUS GGA ⇓ ⇒ PROCESS PARAMETER CONFIGURATION ⇓ EPD PARAMETER ⇓ REFERENCE PARAMETER 646 ⇓ ADJUSTMENT Function description BASIC FUNCTION → PROCESS PARAMETER → ADJUSTMENT ZERO ADJUST This function enables a zero point adjustment to be automatically carried out.
7 Block BASIC FUNCTION Device Functions Proline Promass 83 FOUNDATION Fieldbus Function description BASIC FUNCTION → PROCESS PARAMETER → ADJUSTMENT DENSITY SETPOINT 2 Use this function to enter the density setpoint value for the second fluid for which you (6485) want to carry out field density adjustment.
Device Functions Proline Promass 83 FOUNDATION Fieldbus 7 Block BASIC FUNCTION 7.2.5 Function group PRESSURE CORRECTION ⇒ BASIC FUNCTION FOUNDATION FIELDBUS GGA ⇓ ⇒ PROCESS PARAMETER CONFIGURATION ⇓ EPD PARAMETER ⇓ REFERENCE PARAMETER 646 ⇓ ADJUSTMENT ⇓ PRESSURE CORRECTION 650 Function description BASIC FUNCTION →...
7 Block BASIC FUNCTION Device Functions Proline Promass 83 FOUNDATION Fieldbus Group SYSTEM PARAMETER 7.3.1 Function group CONFIGURATION ⇒ BASIC FUNCTION FOUNDATION FIELDBUS GGA ⇓ PROCESS PARAMETER ⇓ ⇒ SYSTEM PARAMETER CONFIGURATION Function description BASIC FUNCTION → SYSTEM PARAMETER → CONFIGURATION INSTALLATION Use this function to reverse the sign of the flow measured variable, if necessary.
For this reason, the functions described below cannot be changed even when you enter your personal code. Contact the Endress+Hauser service organization if you have any questions about these functions. Note! The individual values of the functions are also provided on the sensor nameplate.
BASIC FUNCTION → SENSOR DATA → FLOW COEFFICIENT All flow coefficients are set at the factory. All the sensor's parameter settings are saved on the S-DAT memory chip. Contact the Endress+Hauser service organization if you have any questions about these functions. TEMPERATURE This function shows the temperature coefficient KM.
BASIC FUNCTION → SENSOR DATA → DENSITY COEFFICIENT All density coefficients are set at the factory. All the sensor's parameter settings are saved on the S-DAT memory chip. Contact the Endress+Hauser service organization if you have any questions about these functions. DENSITY COEFFICIENT This function shows the actual density coefficient C0.
" Caution! These functions are used for displaying device parameters only and consequently cannot be accessed. Contact the Endress+Hauser service organization if you have any questions about these functions. MINIMAL The lowest fluid temperature measured appears on the display. TEMPERATURE...
8 Block SPECIAL FUNCTION Device Functions Proline Promass 83 FOUNDATION Fieldbus Group DENSITY FUNCTIONS 8.1.1 Function group CONFIGURATION ⇒ ⇒ SPECIAL FUNCTION H DENSITY FUNCTIONS CONFIGURATION Function description SPECIAL FUNCTION → DENSITY FUNCTIONS → CONFIGURATION DENSITY FUNCTION Use this function to select the desired density function which is used to calculate special (7000) density values or the percentage proportion of components in two-phase fluids.
Device Functions Proline Promass 83 FOUNDATION Fieldbus 8 Block SPECIAL FUNCTION Function description SPECIAL FUNCTION → DENSITY FUNCTIONS → CONFIGURATION SQUARE EXPANSION Note! COEFFICIENT CARRIER This function is not available unless one of the following was selected in the FLUID...
8 Block SPECIAL FUNCTION Device Functions Proline Promass 83 FOUNDATION Fieldbus Function description SPECIAL FUNCTION → DENSITY FUNCTIONS → CONFIGURATION SQUARE EXPANSION Note! COEFFICIENT TARGET This function is not available unless one of the following was selected in the FLUID...
Device Functions Proline Promass 83 FOUNDATION Fieldbus 8 Block SPECIAL FUNCTION Function description SPECIAL FUNCTION → DENSITY FUNCTIONS → CONFIGURATION MODE Note! (7021) This function is not available unless the FLEXIBLE setting was selected in the DENSITY FUNCTION function (7000).
8 Block SPECIAL FUNCTION Device Functions Proline Promass 83 FOUNDATION Fieldbus Function description SPECIAL FUNCTION → DENSITY FUNCTIONS → CONFIGURATION CONCENTRATION Use this function to select a concentration specification. There are four different SELECTOR concentration specifications available, via which various concentrations can be defined.
Device Functions Proline Promass 83 FOUNDATION Fieldbus 8 Block SPECIAL FUNCTION Function description SPECIAL FUNCTION → DENSITY FUNCTIONS → CONFIGURATION COEFFICIENT A4 Coefficient A4 entry. (7036) User input: 5-digit floating-point number Factory setting: COEFFICIENT B1 Note! (7037) This function only appears if the % MASS 3D or % VOLUME 3D option was selected in the MODE function (7021).
8 Block SPECIAL FUNCTION Device Functions Proline Promass 83 FOUNDATION Fieldbus Group ADVANCED DIAGNOSTICS 8.2.1 Function group CONFIGURATION ⇒ SPECIAL FUNCTION H DENSITY FUNCTIONS ⇓ ⇒ ADVANCED DIAGNOSTICS HEA CONFIGURATION Function description SPECIAL FUNCTION → ADVANCED DIAGNOSTICS → CONFIGURATION REFERENCE Use this function to start determining the user reference status.
Options: PERIODICAL SINGLE SHOT Factory setting: Note! See the Chapter on "Commissioning" in the Operating Instructions Proline Promass 83, BA065D/06/en for more information on advanced diagnostics. ACQUISITION PERIOD Note! (7411) This function is not available unless PERIODICAL was selected in the ACQUISITION MODE function (7410).
8 Block SPECIAL FUNCTION Device Functions Proline Promass 83 FOUNDATION Fieldbus 8.2.3 Function group MASS FLOW ⇒ SPECIAL FUNCTION H DENSITY FUNCTIONS ⇓ ⇒ ADVANCED DIAGNOSTICS HEA CONFIGURATION ⇓ ACQUISITION ⇓ MASS FLOW Function description SPECIAL FUNCTION → ADVANCED DIAGNOSTICS → MASS FLOW Note! The appropriate unit is taken from the UNIT MASS FLOW function (0400).
Device Functions Proline Promass 83 FOUNDATION Fieldbus 8 Block SPECIAL FUNCTION 8.2.4 Function group DENSITY ⇒ SPECIAL FUNCTION H DENSITY FUNCTIONS ⇓ ⇒ ADVANCED DIAGNOSTICS HEA CONFIGURATION ⇓ ACQUISITION ⇓ MASS FLOW ⇓ DENSITY Function description SPECIAL FUNCTION → ADVANCED DIAGNOSTICS → DENSITY Note! The appropriate unit is taken from the UNIT DENSITY function (0420).
8 Block SPECIAL FUNCTION Device Functions Proline Promass 83 FOUNDATION Fieldbus 8.2.5 Function group REFERENCE DENSITY ⇒ SPECIAL FUNCTION H DENSITY FUNCTIONS ⇓ ⇒ ADVANCED DIAGNOSTICS HEA CONFIGURATION ⇓ ACQUISITION ⇓ MASS FLOW ⇓ DENSITY ⇓ REFERENCE DENSITY Function description SPECIAL FUNCTION →...
Device Functions Proline Promass 83 FOUNDATION Fieldbus 8 Block SPECIAL FUNCTION 8.2.6 Function group TEMPERATURE ⇒ SPECIAL FUNCTION H DENSITY FUNCTIONS ⇓ ⇒ ADVANCED DIAGNOSTICS HEA CONFIGURATION ⇓ ACQUISITION ⇓ MASS FLOW ⇓ DENSITY ⇓ REFERENCE DENSITY ⇓ TEMPERATURE Function description SPECIAL FUNCTION →...
8 Block SPECIAL FUNCTION Device Functions Proline Promass 83 FOUNDATION Fieldbus 8.2.7 Function group TUBE DAMPING ⇒ SPECIAL FUNCTION H DENSITY FUNCTIONS ⇓ ⇒ ADVANCED DIAGNOSTICS HEA CONFIGURATION ⇓ ACQUISITION ⇓ MASS FLOW ⇓ DENSITY ⇓ REFERENCE DENSITY ⇓ TEMPERATURE ⇓...
Device Functions Proline Promass 83 FOUNDATION Fieldbus 8 Block SPECIAL FUNCTION 8.2.8 Function group ELECTRODYNAMIC SENSORS ⇒ SPECIAL FUNCTION H DENSITY FUNCTIONS ⇓ ⇒ ADVANCED DIAGNOSTICS HEA CONFIGURATION ⇓ ACQUISITION ⇓ MASS FLOW ⇓ DENSITY ⇓ REFERENCE DENSITY ⇓ TEMPERATURE ⇓...
8 Block SPECIAL FUNCTION Device Functions Proline Promass 83 FOUNDATION Fieldbus Function description SPECIAL FUNCTION → ADVANCED DIAGNOSTICS → ELECTRODYNAMIC SENSORS WARNING LEVEL Note! (7476) This function is not available unless ON was selected in the WARNING MODE function (7403).
Device Functions Proline Promass 83 FOUNDATION Fieldbus 8 Block SPECIAL FUNCTION Function description SPECIAL FUNCTION → ADVANCED DIAGNOSTICS → OPERATING FREQUENCY FLUCTUATION HISTORY The last ten values of the operating frequency fluctuation since the saved values were last OPERATING reset appear on the display.
8 Block SPECIAL FUNCTION Device Functions Proline Promass 83 FOUNDATION Fieldbus Function description SPECIAL FUNCTION → ADVANCED DIAGNOSTICS → TUBE DAMPING FLUCTUATION TUBE DAMPING The measured fluctuation of the tube damping appears on the display. FLUCTUATION Display: (7491) 5-digit floating-point number...
Factory setting: " Caution! • The selection for this function can only be changed by the Endress+Hauser service organization. • If OFF is selected, all parameter changes are not permanently stored in the EEPROM. This means that these parameter changes are no longer available after a power failure.
Device Functions Proline Promass 83 FOUNDATION Fieldbus 9 Block SUPERVISION 9.1.2 Function group OPERATION ⇒ ⇒ SUPERVISION SYSTEM CONFIGURATION ⇓ OPERATION Function description SUPERVISION → SYSTEM → OPERATION ACTUAL Use this function to check the present system status. SYSTEM CONDITION...
9 Block SUPERVISION Device Functions Proline Promass 83 FOUNDATION Fieldbus Function description SUPERVISION → SYSTEM → OPERATION VALUE SIMULATION Note! MEASURAND The function is not visible unless the SIMULATION MEASURAND function (8043) is (8044) active. Use this function to define a selectable value (e.g. 12 m /s).
Device Functions Proline Promass 83 FOUNDATION Fieldbus 9 Block SUPERVISION Group VERSION INFO 9.2.1 Function group DEVICE ⇒ SUPERVISION SYSTEM ⇓ ⇒ VERSION INFO DEVICE Function description SUPERVISION → VERSION INFO → DEVICE DEVICE SOFTWARE Displays the current device software version.
• The language options of the available language group are displayed in the LANGUAGE (2000) function. • You can change the language group via the configuration software ToF Tool-Fieldtool Package. Please do not hesitate to contact your Endress+Hauser sales office if you have any questions. 9.2.4 Function group F-CHIP ⇒...
Device Functions Proline Promass 83 FOUNDATION Fieldbus 9 Block SUPERVISION Function description SUPERVISION → VERSION INFO → F-CHIP SYSTEM OPTION Note! (8241) This function is not available unless the measuring device is equipped an F-CHIP. The software options available in the measuring device appear on the display.
10 Factory settings Device Functions Proline Promass 83 FOUNDATION Fieldbus Factory settings 10.1 SI units (not for USA and Canada) 10.1.1 Low flow cut off, full scale value, pulse value Nominal diameter Low flow cut off Full scale value [mm] (approx.
Device Functions Proline Promass 83 FOUNDATION Fieldbus 10 Factory settings 10.1.3 Density, length, temperature Unit Density kg/l Length Temperature ° C 10.2 US units (only for USA and Canada) 10.2.1 Low flow cut off, full scale value, pulse value Nominal diameter...
Device Functions Proline Promass 83 FOUNDATION Fieldbus 11 Index Function matrix Index Function matrix Blocks 685 = DENSITY COEFFICIENT ....63 686 = ADDITIONAL COEFFICIENT .
Page 94
11 Index Function matrix Device Functions Proline Promass 83 FOUNDATION Fieldbus 1... 6243 = DEVICE REVISION ......51 6244 = DD REVISION .
Page 95
Device Functions Proline Promass 83 FOUNDATION Fieldbus 11 Index Function matrix 7021 = MODE ....... . . 69 7463 = MAXIMUM TUBE DAMPING .
Page 96
11 Index Function matrix Device Functions Proline Promass 83 FOUNDATION Fieldbus Endress+Hauser...
Device Functions Proline Promass 83 FOUNDATION Fieldbus 12 Index local operation Index local operation Flow coefficient ......62 System parameter Access code .
Page 98
12 Index local operation Device Functions Proline Promass 83 FOUNDATION Fieldbus Sensor data ....... . . 61 System .
Page 99
Device Functions Proline Promass 83 FOUNDATION Fieldbus 12 Index local operation Density Advanced diagnostics ..... 75 Gas measurement ....... 19 Device .
Page 100
12 Index local operation Device Functions Proline Promass 83 FOUNDATION Fieldbus Maximal Temperature carrier tube ..... . . 64 Quick Setup ........19 Temperature measured .
Page 101
Device Functions Proline Promass 83 FOUNDATION Fieldbus 12 Index local operation System Configuration ....... 84 Unit Operation .
Page 102
12 Index local operation Device Functions Proline Promass 83 FOUNDATION Fieldbus Endress+Hauser...
Page 103
Device Functions Proline Promass 83 FOUNDATION Fieldbus Table of contents for FOUNDATION Fieldbus Table of contents for FOUNDATION Fieldbus Operation via Discrete Output function block ..188 FOUNDATION Fieldbus..104 Signal processing.
1 Operation via FOUNDATION Fieldbus Device Functions Proline Promass 83 FOUNDATION Fieldbus Operation via FOUNDATION Fieldbus Block model In the FOUNDATION Fieldbus all the device parameters are categorized according to their functional properties and task and are generally assigned to three different blocks. A block may be regarded as a container in which parameters and the associated functionalities are contained.
Page 105
PID block, as a starting variable for connecting the desired application function. Using the Discrete Output function block (DO), various actions and functions in the Device Functions Proline Promass 83 FOUNDATION Fieldbus can be initiated and controlled via FOUNDATION Fieldbus. Note! Additional function blocks such as the PID, Arithmetic, Input Selector, Signal Characterizer and Integrator function block are described in the "FOUNDATION Fieldbus Overview"...
Write protection of the device parameters and simulation in the Analog Input and Discrete Output Function Block can be locked or unlocked via a jumper setting on the FOUNDATION Fieldbus I/O board (→ Operating Instructions, Proline Promass 83 FOUNDATION Fieldbus, BA065D). Endress+Hauser...
Device Functions Proline Promass 83 FOUNDATION Fieldbus 2 Resource Block The parameter WRITE_LOCK shows the status of the hardware write protection. The following statuses are possible: • LOCKED Device data cannot be modified via the FOUNDATION Fieldbus interface. • NOT LOCKED Device data can be modified via the FOUNDATION Fieldbus interface.
2 Resource Block Device Functions Proline Promass 83 FOUNDATION Fieldbus Resource Block parameters The following table shows the Endress+Hauser-specific parameters of the Resource Block. Note! FOUNDATION Fieldbus parameters are described in the Operating Instructions "FOUNDATION Fieldbus Overview" (BA013S) (available at: → www.endress.com → download).
Device Functions Proline Promass 83 FOUNDATION Fieldbus 3 Transducer Block Transducer Block The Transducer Blocks contain all the measurement- and device-specific parameters of the flowmeter. All the settings directly connected with the flow measurement/application are made here. They form the interface between the sensor-specific measured value preprocessing and the Analog Input function blocks required for automation.
3 Transducer Block Device Functions Proline Promass 83 FOUNDATION Fieldbus Signal processing The following figure shows the internal structure of the individual Transducer Blocks: Transducer Block “Flow” Simulation - Value Measurand Simulation - Positive Zero Measurand Return Low Flow Cut Off...
Device Functions Proline Promass 83 FOUNDATION Fieldbus 3 Transducer Block The parameter "Simulation - Value Measurand" ( → Page 129) allows you to specify a simulation value for the Transducer Block in order to test assigned parameters in the device and subsequent function blocks.
In addition, the active device error is displayed via the "Diagnosis" Transducer Block in the "Diag. - Act.Sys.Condition" parameter ( → Page 131). For more information on rectifying errors → See Operating Instructions for Proline Promass 83 FOUNDATION Fieldbus (BA065D), "Troubleshooting" section.
Detailed information on the current device status is displayed via the "Diagnosis" Transducer Block in the manufacturer-specific parameter "Diag. - Act.Sys.Condition" ( → Page 131). For more information on rectifying errors → See Operating Instructions for Proline Promass 83 FOUNDATION Fieldbus (BA065D), "Troubleshooting" section.
Page 114
• You can disable programming again by entering any number (other than the release code) in this parameter. • The Endress+Hauser service organization can be of assistance if you mislay your personal code. • Certain parameters are not accessible unless a special service code is entered.
Page 115
Device Functions Proline Promass 83 FOUNDATION Fieldbus 3 Transducer Block "Flow" Transducer Block / base index 1400: Write access with operating Parameter Description mode (MODE_BLK) System Value - read only Use this parameter to view the volume flow.The volume flow is derived Volume Flow from the measured mass flow and the measured density of the fluid.
Page 116
3 Transducer Block Device Functions Proline Promass 83 FOUNDATION Fieldbus "Flow" Transducer Block / base index 1400: Write access with operating Parameter Description mode (MODE_BLK) System Unit - AUTO - OOS Use this parameter to select the unit for the corrected volume flow Corr.Volume Flow...
Page 117
Device Functions Proline Promass 83 FOUNDATION Fieldbus 3 Transducer Block "Flow" Transducer Block / base index 1400: Write access with operating Parameter Description mode (MODE_BLK) System Unit - AUTO - OOS Use this parameter to select the unit for the reference density.
Page 118
3 Transducer Block Device Functions Proline Promass 83 FOUNDATION Fieldbus "Flow" Transducer Block / base index 1400: Write access with operating Parameter Description mode (MODE_BLK) System Unit - AUTO - OOS Use this parameter to select the unit for pressure.
Page 119
Device Functions Proline Promass 83 FOUNDATION Fieldbus 3 Transducer Block "Flow" Transducer Block / base index 1400: Write access with operating Parameter Description mode (MODE_BLK) Low Flow Cut Off - AUTO - OOS Use this function to enter the off-value (b) for low flow cut off.
Page 120
3 Transducer Block Device Functions Proline Promass 83 FOUNDATION Fieldbus "Flow" Transducer Block / base index 1400: Write access with operating Parameter Description mode (MODE_BLK) Process - Pressure AUTO - OOS Activation of the pressure shock suppression Shock Suppression Pressure shock suppression is activated after the flow falls below the (continued) ON-VALUE for the low flow cut off (see point a in graphic).
Page 121
Device Functions Proline Promass 83 FOUNDATION Fieldbus 3 Transducer Block "Flow" Transducer Block / base index 1400: Write access with operating Parameter Description mode (MODE_BLK) Sys. - Permanent read only This parameter indicates whether automatic, permanent storage of Storage parameter changes in the EEPROM is switched on or off.
Page 122
3 Transducer Block Device Functions Proline Promass 83 FOUNDATION Fieldbus "Flow" Transducer Block / base index 1400: Write access with operating Parameter Description mode (MODE_BLK) EPD - AUTO - OOS Use this parameter to enter the time span for which the criteria for an empty...
Page 123
Device Functions Proline Promass 83 FOUNDATION Fieldbus 3 Transducer Block "Flow" Transducer Block / base index 1400: Write access with operating Parameter Description mode (MODE_BLK) Ref.Param. - AUTO - OOS Note! Exp. Coeff.Lin. The entry is not active unless the CALC. REF. DENSITY option was selected in the parameter "Reference Param.
Page 124
3 Transducer Block Device Functions Proline Promass 83 FOUNDATION Fieldbus "Flow" Transducer Block / base index 1400: Write access with operating Parameter Description mode (MODE_BLK) Adj. - Zeropoint AUTO - OOS This parameter enables a zero point adjustment to be automatically carried Adjustment out.
Page 125
Device Functions Proline Promass 83 FOUNDATION Fieldbus 3 Transducer Block "Flow" Transducer Block / base index 1400: Write access with operating Parameter Description mode (MODE_BLK) Adj. - Dens. AUTO - OOS With this parameter a density adjustment can be carried out. The density Adjustment adjustment values will thus be recalculated and stored.
Page 126
3 Transducer Block Device Functions Proline Promass 83 FOUNDATION Fieldbus "Flow" Transducer Block / base index 1400: Write access with operating Parameter Description mode (MODE_BLK) Press.Corr. - AUTO - OOS Note! Pressure The entry is not active unless the FIX option was selected in the "Press.Corr.
Page 127
Device Functions Proline Promass 83 FOUNDATION Fieldbus 3 Transducer Block "Flow" Transducer Block / base index 1400: Write access with operating Parameter Description mode (MODE_BLK) Sys. - Positive Zero AUTO - OOS Use this parameter to interrupt evaluation of measured variables. This is Return necessary when a piping system is being cleaned, for example.
Page 128
3 Transducer Block Device Functions Proline Promass 83 FOUNDATION Fieldbus "Flow" Transducer Block / base index 1400: Write access with operating Parameter Description mode (MODE_BLK) Sensor Data - read only This function shows the actual density coefficient C0. Density Coeff.
Page 129
Device Functions Proline Promass 83 FOUNDATION Fieldbus 3 Transducer Block "Flow" Transducer Block / base index 1400: Write access with operating Parameter Description mode (MODE_BLK) Sensor Data - Min. read only The lowest carrier tube temperature measured appears on the display.
Page 130
3 Transducer Block Device Functions Proline Promass 83 FOUNDATION Fieldbus "Flow" Transducer Block / base index 1400: Write access with operating Parameter Description mode (MODE_BLK) Simulation - read only Displays the unit used for the simulation. Unit Note! Depending on the selection in the parameter "Simulation - Measurand", the unit is taken over from the corresponding parameter "System Unit -".
Device Functions Proline Promass 83 FOUNDATION Fieldbus 3 Transducer Block "Diagnosis" Transducer Block parameters The following table shows the Endress+Hauser-specific parameters of the "Diagnosis" Transducer Block. These can only be changed after entering a release code in the "Access - Code" parameter. Note! FOUNDATION Fieldbus parameters are described in the Operating Instructions "FOUNDATION...
Page 132
3 Transducer Block Device Functions Proline Promass 83 FOUNDATION Fieldbus "Diagnosis" Transducer Block/base index 1600 Write access with operating Parameter Description mode (MODE_BLK) Sys. - Alarm Delay AUTO - OOS Use this function to define a time span for which the criteria for an error have to be satisfied without interruption before a fault or notice message is generated.
Page 133
Device Functions Proline Promass 83 FOUNDATION Fieldbus 3 Transducer Block "Diagnosis" Transducer Block/base index 1600 Write access with operating Parameter Description mode (MODE_BLK) Sys. - AUTO - OOS Use this parameter to save the parameter settings / configuration of the...
Device Functions Proline Promass 83 FOUNDATION Fieldbus "Display" Transducer Block parameters The following table shows the Endress+Hauser-specific parameters of the "Display" Transducer Block. These can only be changed after entering a release code in the "Access - Code" parameter. Note! FOUNDATION Fieldbus parameters are described in the Operating Instructions "FOUNDATION...
Page 135
Device Functions Proline Promass 83 FOUNDATION Fieldbus 3 Transducer Block "Display" Transducer Block/base index 1800 Write access with operating Parameter Description mode (MODE_BLK) Access - AUTO - OOS Use this function to enter a personal code to enable configuration. This Def.Private Code...
Page 136
3 Transducer Block Device Functions Proline Promass 83 FOUNDATION Fieldbus "Display" Transducer Block/base index 1800 Write access with operating Parameter Description mode (MODE_BLK) Config. - AUTO - OOS Use this parameter to enter a time constant defining how the display reacts...
Page 137
Device Functions Proline Promass 83 FOUNDATION Fieldbus 3 Transducer Block "Display" Transducer Block/base index 1800 Write access with operating Parameter Description mode (MODE_BLK) 1 = Main line 2 = Additional line 3 = Info line A0001253 Main Line - AUTO - OOS...
Page 138
3 Transducer Block Device Functions Proline Promass 83 FOUNDATION Fieldbus "Display" Transducer Block/base index 1800 Write access with operating Parameter Description mode (MODE_BLK) Main Line - Advanced options with optional software package Assign ADVANCED DIAGNOSTICS: (continued) MASS FLOW DEVIATION DENSITY DEVIATION...
Page 139
Device Functions Proline Promass 83 FOUNDATION Fieldbus 3 Transducer Block "Display" Transducer Block/base index 1800 Write access with operating Parameter Description mode (MODE_BLK) Main Line Mux - AUTO - OOS Use this parameter to define the second reading to be displayed in the main...
Page 140
3 Transducer Block Device Functions Proline Promass 83 FOUNDATION Fieldbus "Display" Transducer Block/base index 1800 Write access with operating Parameter Description mode (MODE_BLK) Main Line Mux - AUTO - OOS Note! 100%-Value The entry is not active unless one of the following was selected in the parameter "Main Line Mux - Assign":...
Page 141
Device Functions Proline Promass 83 FOUNDATION Fieldbus 3 Transducer Block "Display" Transducer Block/base index 1800 Write access with operating Parameter Description mode (MODE_BLK) 1 = Main line 2 = Additional line 3 = Info line A0001253 Add. Line - AUTO - OOS...
Page 142
3 Transducer Block Device Functions Proline Promass 83 FOUNDATION Fieldbus "Display" Transducer Block/base index 1800 Write access with operating Parameter Description mode (MODE_BLK) Add. Line - Advanced options with optional software package Assign ADVANCED DIAGNOSTICS: (continued) MASS FLOW DEVIATION DENSITY DEVIATION...
Page 143
Device Functions Proline Promass 83 FOUNDATION Fieldbus 3 Transducer Block "Display" Transducer Block/base index 1800 Write access with operating Parameter Description mode (MODE_BLK) Add. Line - AUTO - OOS Note! Display Mode The option is not active unless one of the following was selected in the parameter "Add.
Page 144
3 Transducer Block Device Functions Proline Promass 83 FOUNDATION Fieldbus "Display" Transducer Block/base index 1800 Write access with operating Parameter Description mode (MODE_BLK) Add. Line Mux - Advanced options with optional software package Assign CONCENTRATION: (continued) TARGET MASS FLOW % TARGET MASS FLOW...
Page 145
Device Functions Proline Promass 83 FOUNDATION Fieldbus 3 Transducer Block "Display" Transducer Block/base index 1800 Write access with operating Parameter Description mode (MODE_BLK) Add. Line Mux - AUTO - OOS Note! Format The option is not active unless a number was selected in the parameter "Add.
Page 146
3 Transducer Block Device Functions Proline Promass 83 FOUNDATION Fieldbus "Display" Transducer Block/base index 1800 Write access with operating Parameter Description mode (MODE_BLK) 1 = Main line 2 = Additional line 3 = Info line A0001253 Info Line - AUTO - OOS...
Page 147
Device Functions Proline Promass 83 FOUNDATION Fieldbus 3 Transducer Block "Display" Transducer Block/base index 1800 Write access with operating Parameter Description mode (MODE_BLK) Info Line - Advanced options with optional software package Assign ADVANCED DIAGNOSTICS: (continued) MASS FLOW DEVIATION DENSITY DEVIATION...
Page 148
3 Transducer Block Device Functions Proline Promass 83 FOUNDATION Fieldbus "Display" Transducer Block/base index 1800 Write access with operating Parameter Description mode (MODE_BLK) Info Line - AUTO - OOS Note! Display Mode The option is not active unless one of the following was selected in the parameter "Info Line - Assign"...
Page 149
Device Functions Proline Promass 83 FOUNDATION Fieldbus 3 Transducer Block "Display" Transducer Block/base index 1800 Write access with operating Parameter Description mode (MODE_BLK) Info Line Mux - Advanced options with optional software package Assign CONCENTRATION: (continued) TARGET MASS FLOW % TARGET MASS FLOW...
Page 150
3 Transducer Block Device Functions Proline Promass 83 FOUNDATION Fieldbus "Display" Transducer Block/base index 1800 Write access with operating Parameter Description mode (MODE_BLK) Info Line Mux - AUTO - OOS Note! Format The option is not active unless a number was selected in the parameter "Info Line Mux - Assign"...
Device Functions Proline Promass 83 FOUNDATION Fieldbus 3 Transducer Block "Totalizer" Transducer Block parameters The following table shows the Endress+Hauser-specific parameters of the "Totalizer" Transducer Block. These can only be changed after entering a release code in the "Access - Code" parameter. Note! FOUNDATION Fieldbus parameters are described in the Operating Instructions "FOUNDATION...
Page 152
3 Transducer Block Device Functions Proline Promass 83 FOUNDATION Fieldbus "Totalizer" Transducer Block/base index 1900 Write access with operating Parameter Description mode (MODE_BLK) Tot. 1 to 3 - Sum AUTO - OOS Use this parameter to view the total for the totalizer’s measured variable aggregated since measuring commenced.
Page 153
Device Functions Proline Promass 83 FOUNDATION Fieldbus 3 Transducer Block "Totalizer" Transducer Block/base index 1900 Write access with operating Parameter Description mode (MODE_BLK) Tot. 1 to 3 - Assign AUTO - OOS Use this parameter to assign a measured variable to the totalizer.
Page 154
3 Transducer Block Device Functions Proline Promass 83 FOUNDATION Fieldbus "Totalizer" Transducer Block/base index 1900 Write access with operating Parameter Description mode (MODE_BLK) Tot. - Reset All AUTO - OOS Reset all totalizers simultaneously to zero ("Tot. 1 to 3 - Sum" parameter).
Page 155
Otherwise the value "NaN" (not-a-number) or "Not licensed" is displayed in a parameter. The following table shows the Endress+Hauser-specific parameters of the "Calculated Density" Transducer Block. These can only be changed after entering a release code in the "Access - Code"...
Page 156
3 Transducer Block Device Functions Proline Promass 83 FOUNDATION Fieldbus "Calculated Density" Transducer Block/base index 2000 Write access with operating Parameter Description mode (MODE_BLK) System Value - read only Note! Target Mass Flow The option is not active unless one of the following was selected: •...
Page 157
Device Functions Proline Promass 83 FOUNDATION Fieldbus 3 Transducer Block "Calculated Density" Transducer Block/base index 2000 Write access with operating Parameter Description mode (MODE_BLK) System Value - read only Note! Perc.Target Volume The option is not active unless one of the following was selected: Flow •...
Page 158
3 Transducer Block Device Functions Proline Promass 83 FOUNDATION Fieldbus "Calculated Density" Transducer Block/base index 2000 Write access with operating Parameter Description mode (MODE_BLK) System Value - read only Note! Carrier Mass Flow The option is not active unless one of the following was selected: •...
Page 159
Device Functions Proline Promass 83 FOUNDATION Fieldbus 3 Transducer Block "Calculated Density" Transducer Block/base index 2000 Write access with operating Parameter Description mode (MODE_BLK) System Value - read only Note! Perc.Carrier The option is not active unless one of the following was selected: Volume Flow •...
Page 160
3 Transducer Block Device Functions Proline Promass 83 FOUNDATION Fieldbus "Calculated Density" Transducer Block/base index 2000 Write access with operating Parameter Description mode (MODE_BLK) Dens. - Function AUTO - OOS Use this parameter to select the desired density function which is used to calculate special density values or the percentage proportion of components in two-phase fluids.
Page 161
Device Functions Proline Promass 83 FOUNDATION Fieldbus 3 Transducer Block "Calculated Density" Transducer Block/base index 2000 Write access with operating Parameter Description mode (MODE_BLK) Dens. - AUTO - OOS Note! Sqr.Exp.Coef. The option is not active unless one of the following was selected in the Carrier parameter "Dens.
Page 162
3 Transducer Block Device Functions Proline Promass 83 FOUNDATION Fieldbus "Calculated Density" Transducer Block/base index 2000 Write access with operating Parameter Description mode (MODE_BLK) Dens. - AUTO - OOS Note! Sqr.Exp.Coef.Target The option is not active unless one of the following was selected in the parameter "Dens.
Page 163
Device Functions Proline Promass 83 FOUNDATION Fieldbus 3 Transducer Block "Calculated Density" Transducer Block/base index 2000 Write access with operating Parameter Description mode (MODE_BLK) Dens. - AUTO - OOS Note! Ref.Temperature The option is only active if the OFF, °BRIX or FLEXIBLE option was not selected in the "Dens.
Page 164
3 Transducer Block Device Functions Proline Promass 83 FOUNDATION Fieldbus "Calculated Density" Transducer Block/base index 2000 Write access with operating Parameter Description mode (MODE_BLK) Dens. - AUTO - OOS Use this parameter to select a concentration specification. There are four...
Page 165
Device Functions Proline Promass 83 FOUNDATION Fieldbus 3 Transducer Block "Calculated Density" Transducer Block/base index 2000 Write access with operating Parameter Description mode (MODE_BLK) Dens. - A3 read only Coefficient A3 entry. User input: 5-digit floating-point number Factory setting: Dens. - A4 read only Coefficient A4 entry.
(not-a-number) or "Not licensed" is displayed in a parameter. • This order option is only available for Proline Promass 83I. The following table shows the Endress+Hauser-specific parameters of the "Viscosity" Transducer Block. These can only be changed after entering a release code in the "Access - Code" parameter.
Page 167
Device Functions Proline Promass 83 FOUNDATION Fieldbus 3 Transducer Block "Viscosity" Transducer Block/base index 2100 Write access with operating Parameter Description mode (MODE_BLK) Visc. - Kin.Viscosity read only Use this function to view the calculated kinematic viscosity. Display: 5-digit floating-point number, incl. unit.
Page 168
3 Transducer Block Device Functions Proline Promass 83 FOUNDATION Fieldbus "Viscosity" Transducer Block/base index 2100 Write access with operating Parameter Description mode (MODE_BLK) Visc. - X1 AUTO - OOS Use this function to enter the compensation coefficient X1 (calculation → See Special Documentation on viscosity measurement for Proline Promass 83I, SD102D).
Otherwise the value "NaN" (not-a-number) or "Not licensed" is displayed in a parameter. The following table shows the Endress+Hauser-specific parameters of the "Advanced Diagnostics" Transducer Block. These can only be changed after entering a release code in the "Access - Code"...
Page 170
3 Transducer Block Device Functions Proline Promass 83 FOUNDATION Fieldbus "Advanced Diagnostics" Transducer Block/base index 2200 Write access with operating Parameter Description mode (MODE_BLK) Adv. - AUTO - OOS Use this parameter to start determining the user reference status. The Ref.Cond.User...
Page 171
Device Functions Proline Promass 83 FOUNDATION Fieldbus 3 Transducer Block "Advanced Diagnostics" Transducer Block/base index 2200 Write access with operating Parameter Description mode (MODE_BLK) Adv. - AUTO - OOS Use this parameter to specify whether the advanced diagnostics parameters Acquisition Mode should be determined on a periodical or single-shot basis.
Page 172
3 Transducer Block Device Functions Proline Promass 83 FOUNDATION Fieldbus "Advanced Diagnostics" Transducer Block/base index 2200 Write access with operating Parameter Description mode (MODE_BLK) System Unit - AUTO - OOS Use this function to select the unit for displaying the mass.
Page 173
Device Functions Proline Promass 83 FOUNDATION Fieldbus 3 Transducer Block "Advanced Diagnostics" Transducer Block/base index 2200 Write access with operating Parameter Description mode (MODE_BLK) Density - read only The reference value for the density appears on the display. Ref.Value Factory Display: 5-digit floating-point number, incl.
Page 174
3 Transducer Block Device Functions Proline Promass 83 FOUNDATION Fieldbus "Advanced Diagnostics" Transducer Block/base index 2200 Write access with operating Parameter Description mode (MODE_BLK) Density - Warning AUTO - OOS Note! Level This option is not active unless ON was selected in the parameter "Adv.
Page 175
Device Functions Proline Promass 83 FOUNDATION Fieldbus 3 Transducer Block "Advanced Diagnostics" Transducer Block/base index 2200 Write access with operating Parameter Description mode (MODE_BLK) Ref.Density - AUTO - OOS Note! Warning Level This option is not active unless ON was selected in the parameter "Adv.
Page 176
3 Transducer Block Device Functions Proline Promass 83 FOUNDATION Fieldbus "Advanced Diagnostics" Transducer Block/base index 2200 Write access with operating Parameter Description mode (MODE_BLK) Temperature - read only This function displays the deviation between the measured temperature and Deviation the reference values selected in the parameter "Adv. - Select Ref.Condition"...
Page 177
Device Functions Proline Promass 83 FOUNDATION Fieldbus 3 Transducer Block "Advanced Diagnostics" Transducer Block/base index 2200 Write access with operating Parameter Description mode (MODE_BLK) Tube Damping - AUTO - OOS Note! Warning Level This option is not active unless ON was selected in the parameter "Adv.
Page 178
3 Transducer Block Device Functions Proline Promass 83 FOUNDATION Fieldbus "Advanced Diagnostics" Transducer Block/base index 2200 Write access with operating Parameter Description mode (MODE_BLK) El.Dyn.Sens. - AUTO - OOS Note! Warning Level This option is not active unless ON was selected in the parameter "Adv.
Page 179
Device Functions Proline Promass 83 FOUNDATION Fieldbus 3 Transducer Block "Advanced Diagnostics" Transducer Block/base index 2200 Write access with operating Parameter Description mode (MODE_BLK) Tube Damp.Fluct. - read only The reference value for the fluctuation of the tube damping appears on the Ref.Value Factory...
4 Function blocks Device Functions Proline Promass 83 FOUNDATION Fieldbus Function blocks The function blocks contain the basic automation functions of the field device. We distinguish between different function blocks, e.g. Analog Input function block, PID function block (PID controller), etc.
Device Functions Proline Promass 83 FOUNDATION Fieldbus 5 Analog Input function block Analog Input function block In the Analog Input function block (AI) the process variables from the Transducer Block are prepared for the subsequent automation functions (e.g. scaling, limit value processing). The automation function is defined by the connections of the outputs.
Page 182
The simulation mode is activated by means of jumpers on the I/O board (→ See Operating Instructions for Proline Promass 83 FOUNDATION Fieldbus, BA065D). The parameter L_TYPE is used to select the linearization type of the input or simulation value: •...
Device Functions Proline Promass 83 FOUNDATION Fieldbus 5 Analog Input function block Important functions and parameters of the Analog Input function blocks The most important functions and parameters of the Analog Input function blocks are listed below. Note! All the FOUNDATION Fieldbus parameters available are described in the Operating Instructions "FOUNDATION Fieldbus Overview"...
• Detailed information on the current device status is displayed via the "Diagnosis" Transducer Block in the parameter "Diag. - Act.Sys.Condition". A list of all the error messages, including remedial measures, can be found in the Operating Instructions for Proline Promass 83 FOUNDATION Fieldbus (BA065D).
Detailed information on the current device status is displayed via the "Diagnosis" Transducer Block in the manufacturer-specific parameter "Diag. - Act.Sys.Condition" ( → Page 131). For more information on rectifying errors → See Operating Instructions for Proline Promass 83 FOUNDATION Fieldbus (BA065D), "Troubleshooting" section.
5 Analog Input function block Device Functions Proline Promass 83 FOUNDATION Fieldbus • Parameter group OUT_SCALE OUT_SCALE 0 % OUT_SCALE 100 % OUT_SCALE UNIT The result is that with an input value of, for example, 15 kg/h a value of 50% is output via the parameter OUT.
Page 187
Device Functions Proline Promass 83 FOUNDATION Fieldbus 5 Analog Input function block Limit value process alarms If a limit value is infringed then the priority specified for the limit value alarm will be checked before the limit value violation is communicated to the fieldbus host system. The priority that specifies the action in the event of an active limit value violation is determined by the following parameters: •...
6 Discrete Output function block Device Functions Proline Promass 83 FOUNDATION Fieldbus Discrete Output function block The Discrete Output function block (DO, Discrete Output) processes a discrete setpoint value received from an upstream function block or higher level process control system, with which various instrument functions (e.g.
Device Functions Proline Promass 83 FOUNDATION Fieldbus 6 Discrete Output function block of the Discrete Output function block is communicated to the process control system as an answer message via parameter RCAS_OUT_D. In the AUTO operating mode (automatic operation), the set point value (parameter SP_D) is prescribed directly in the Discrete Output function block.
6 Discrete Output function block Device Functions Proline Promass 83 FOUNDATION Fieldbus 6.2.4 Values for the parameters CAS_IN_D, RCAS_IN_D, OUT_D, and SP_D Via the Discrete Output function block, different instrument functions in the Transducer Block can be initiated via manufacturer-specific, fixed set point values from an upstream function block.
Page 191
Device Functions Proline Promass 83 FOUNDATION Fieldbus 6 Discrete Output function block Example for controlling positive zero return via the Discrete Output function block. The following example shows how positive zero return can be activated or deactivated via the Discrete Output function block during a cleaning procedure.
7 Additional function blocks Device Functions Proline Promass 83 FOUNDATION Fieldbus Additional function blocks Note! Additional function blocks such as the PID, Arithmetic, Input Selector, Signal Characterizer and Integrator function block are described in the "FOUNDATION Fieldbus Overview" (BA013S) Operating Instructions (available at: → www.endress.com → download).
Device Functions Proline Promass 83 FOUNDATION Fieldbus 8 Factory settings Factory settings SI units (not for USA and Canada) 8.1.1 Low flow cut off, full scale value, pulse value Nominal diameter Low flow cut off Full scale value [mm] (approx. v = 0.04 m/s) (approx.
8 Factory settings Device Functions Proline Promass 83 FOUNDATION Fieldbus 8.1.3 Density, length, temperature Unit Density kg/l Length Temperature ° C US units (only for USA and Canada) 8.2.1 Low flow cut off, full scale value, pulse value Nominal diameter...
Device Functions Proline Promass 83 FOUNDATION Fieldbus 9 Index FOUNDATION Fieldbus Index FOUNDATION Fieldbus Block status Resource Block ......106 Access Transducer Block .
Page 196
9 Index FOUNDATION Fieldbus Device Functions Proline Promass 83 FOUNDATION Fieldbus Fault message Parameters See Operating Instructions for Resource Block ......108 Proline Promass 83 FOUNDATION Fieldbus (BA065D) Pressure shock suppression .
Page 197
Device Functions Proline Promass 83 FOUNDATION Fieldbus 9 Index FOUNDATION Fieldbus Sys. - Sim. Failsafe Mode Totalizer "Diagnosis" Transducer Block ....132 Define failsafe mode ......154 Sys.