26
Specific absorption rates
The specific absorption rate for the RF power and the rate of change of the gradient fields
are checked according to the requirements of norm IEC 60601-2-33.
Cooling system
Two different customer specific cooling alternatives (Separator or Eco Chiller) are avail-
able.
Separator option (for connec-
tion to available cooling sys-
tem)
Eco Chiller option with auto-
matic adaptation to the
required cooling demands (e.g.
different night/day mode) to
decrease energy cost
1. Water temperature: 12 °C (45 °F)
2. Free Cooling Unit, optional
3. In case of clinical routine measurement conditions
Siemens AG Healthcare Sector
Water consumption
Heat dissipation to water
GREEN Cooling Package
•
Automatic start if the surrounding temperature is
18 °C (64 °F) or less
•
If the temperature is less than –10 °C (14 °F) the
chiller is switched off
11/12
90 l/min
60 kW
2
:
3
1
Technical data