Guidance and manufacturer's declaration — electromagnetic
immunity
The MR system is intended for use in the electromagnetic environment specified below.
The customer or the user of the MR system should assure that it is used in such an environ-
ment.
Immunity tests
Electrostatic dis-
charge (ESD)
IEC 61000-4-2
Electrical fast tran-
sient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Siemens AG Healthcare Sector
IEC 60601 test level
±6 kV contact
±8 kV air
±2 kV for power sup-
ply lines
±1 kV for signal lines
±1 kV differential
mode
±2 kV common mode
Compliance level
±6 kV contact
±8 kV air
±2 kV for power sup-
ply lines
±1 kV for signal lines
±1 kV differential
mode
±2 kV common mode
11/12
Guidance and manufacturer's declaration EMC
Electromagnetic environ-
ment — guidance
Floors should be wood, con-
crete or ceramic tile.
If floors are covered with syn-
thetic material, the relative
humidity should be at least
30%.
Mains power quality should be
that of a typical commercial or
hospital environment.
Mains power quality should be
that of a typical commercial or
hospital environment.