Siemens MAGNETOM Skyra Owner's Manual page 99

Mr system
Hide thumbs Also See for MAGNETOM Skyra:
Table of Contents

Advertisement

Line power supply
XQ gradients
Voltage
Frequency
Connection value
On-site protection
For all products, line power has to be supplied via an on-site system contact or via another
multipole shut-down mechanism. Room installation has to be in compliance with
VDE 0100-710
1. In all countries, compliance with local and national legal regulations is required.
However, we strongly recommend compliance with the regulations described herein – to the extent per-
mitted by relevant local and national laws – in order to ensure the safety of operating personnel, patients,
and third parties.
Power rating plate
Siemens AG Healthcare Sector
Values
380 V, 400 V, 420 V, 440 V,
460 V, 480 V
50/60 Hz
110 kVA
160 A
1
.
Tolerance
±10%
±1 Hz
11/12
27
Technical data