RF data
Peak power of transmitter amplifier
Channel 0
Channel 1
Transmitter bandwidth
Receiver bandwidth
RF transmit coils
Body coil
CP Extremity coil
Tx/Rx 15-Channel Knee
coil
Tx/Rx Knee 15 MR coil
Tx/Rx CP Head coil
Siemens AG Healthcare Sector
40.0 kW
≥16.0 KW
≥24.0 KW
800 kHz
500 Hz–1 MHz (for each channel)
Max. applied RF field B1+
Max. specified B1+ rms
Distance to isocenter, where RF transmit field is reduced by
3 dB
10 dB
Max. applied RF field B1+
Max. specified B1+ rms
Max. applied RF field B1+
Max. specified B1+ rms
Max. applied RF field B1+
Max. specified B1+ rms
Max. applied RF field B1+
Max. specified B1+ rms
11/12
24 µT
3.6 µT
0.15 m
0.27 m
47 µT
9 µT
47 µT
8 µT
47 µT
8 µT
41.1 µT
6 µT
MR compatibility data sheet