MR compatibility protocols
According to IEC 60601-2-33, protocols need to be proposed for testing the functionality
of non-Siemens MR products. The protocols listed below are routinely used in the system.
!
The tests are not used to evaluate the effects of the non-Siemens MR product on the
image quality of the MR system.
!
Successful tests with the below protocols do not guarantee MR compatibility of
non-Siemens MR products.
Directory
head/library/T2
head/library/SPACE
abdomen/library/T1
abdomen/library/T2
abdomen/library/3D
Siemens AG Healthcare Sector
Protocol
t2_tse_tra_320_p2
t2_spc_sag_p2_iso
t1_fl2d_fs_tra_mbh_320
t2_tse_tra_fs_p2_mbh_320
t1_vibe_fs_tra_bh_p2
11/12
3
MR compatibility data sheet