Siemens MAGNETOM Skyra Owner's Manual page 65

Mr system
Hide thumbs Also See for MAGNETOM Skyra:
Table of Contents

Advertisement

Guidance and manufacturer's declaration — electromagnetic
emissions
The MR system is intended for use in the electromagnetic environment specified below.
The customer or the user of the MR system should assure that it is used in such an environ-
ment.
Emissions test
Radiated RF emissions
CISPR 11
Conducted RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Note: It is essential that the actual shielding effectiveness and filter attenuation of the shielded location
be verified to assure that they meet the minimum specification.
Siemens AG Healthcare Sector
Compliance
Class A/Group 2
Class A/Group 2
Not applicable
Not applicable
Electromagnetic environment — guidance
The MR system must emit electromagnetic energy
in order to perform its intended function.
Nearby electronic equipment may be affected.
The MR system must be used only in a shielded
location with a minimum RF shielding effective-
ness and, for each cable that exits the shielded
location, a minimum RF filter attenuation of 80 dB
from 10 MHz to 20 MHz, 100 dB from 20 MHz to
80 MHz and 80 dB from 80 MHz to 100 MHz. (The
minimum at 20 MHz is 100 dB and the minimum
at 80 MHz is 80 dB.)
The MR system, when installed in such a shielded
location, is suitable for use in all establishments
other than domestic and those directly connected
to the public low-voltage power supply network
that supplies buildings used for domestic pur-
poses.
11/12
Guidance and manufacturer's declaration EMC
3