Operator Responsibilities for Maintenance
Preventive maintenance
The purpose of preventive maintenance is to keep unforeseen failures to a minimum. This
satisfies the prerequisites for system compliance with the guaranteed, long-term charac-
teristics.
The effects of different operating conditions (full or partial load operation, temperature,
size of dust particles, humidity, gases, vapors) are checked and the condition of parts sub-
ject to wear is determined by recording and analyzing characteristic values. Preventive
measures must be adopted or repairs must be undertaken as appropriate.
The specified maintenance intervals correspond to the minimum requirements. Compli-
ance with stricter national legislation may occasionally be necessary.
Object or function
Cooling system
Air filters, fans
Chiller (option)
© Siemens, 2009
Reason
Preventive measure to
avoid overheating
Preventive measure to
avoid pollution
Preventive measure to
avoid overheating
M7-000.664.01.04.02
03.12
What is checked:
Check for leaks and condensation
Check/replace water filter
Check water pressure and add water if
necessary
Replace air filter
Check functionality of fans
1
Page 11 of 14
CS SD MR
11
Interval
Annually
Annually
1