12
Object or function
LCD color display
Phantom
Examination room door
Magnet
Helium compressor
Adsorber
Patient table
MAGNETOM
Operator Responsibilities for Maintenance
Reason
Optimum function
within specifications
Safety of patient, per-
sonnel, system, and
environment
Preventive measure to
avoid wear and tear
Optimum function
within specifications
Safety of patient, per-
sonnel, and system
Optimum function
within specifications
Optimum function
within specifications
Preventive measure to
avoid wear and tear
M7-000.664.01.04.02
03.12
What is checked:
Check/adjust according to specifica-
tions
Phantoms cannot have any defects or
air bubbles. Replace or refill as neces-
sary.
2
Check/set pressure on magnet and
compressor
LHe Refill
4
Check/Replace
the cold head
Check for leaks
Replace adsorber
Check the hydraulic system
Check/refill hydraulic oil
Page 12 of 14
CS SD MR
Interval
Annually
Annually
2
Annually
3
n.a.
n.a.
Annually
Every
3 years
Annually