TimTX TrueForm
Innovative techniques in the RF excitation hardware
Uniform RF distribution in all body regions
TrueForm excitation uses amplitude and phase transmission settings optimized for
dedicated body regions. Feeding the 2 ports of the integrated body coil with an opti-
mized weighting yields a homogeneous B1 distribution
a-SPACE
B1 Filter
RF receiver technology
Number of coil elements
Number of independent receiver chan-
nels
Quadrature demodulation and filtering Digital
Receiver bandwidth
Receiver signal resolution
ADC sampling rate
Preamplifier noise figure
Dynamic range at coil connector
(referred to 1 Hz resolution bandwidth)
1. Optional
Siemens AG Healthcare Sector
A version of the SPACE sequence. a-SPACE uses composite
adiabatic excitation pulses, which are insensitive to B1
spatial variations
An adaptive inline image filter that reduces any remnant
B1 effects without affecting image contrast
Up to 204
48, 64
500 Hz–1 MHz (for each channel)
32 bit
80 MHz
<0.5 dB
164 dB instantaneous at receiver
169 dB with automatic gain control at local
coil connector
11/12
1
1
, 128
Technical data