A MR system is classified as a large and permanently-installed equipment. The exemption
according to the requirement 36.202.3 b) 9. of IEC 60601-1-2 Edition 2.1 (2004-11) and
6.2.3.2i of IEC 60601-1-2 Edition 3 (2007-03) has been used and the equipment was not
tested for radiated RF immunity over the entire frequency range 80 MHz to 2.5 GHz.
!
The MR system has been tested for radiated RF immunity only at selected frequencies.
Tested ISM frequencies
Modulation characteristic
Test level
Signal generator with amplifier
Siemens AG Healthcare Sector
80 MHz–2.5 GHz
1 kHz, 80% AM
3 V/m
9 kHz to 3.2 GHz
11/12
Guidance and manufacturer's declaration EMC