8
Object or function
Tim dockable table
(option)
Intercom, pneumatic
bulb
Monitoring the magnet
(MSUP)
Magnet stop function
(ERDU)
Gradient supervision
QA (Quality Assurance)
measurements
1. According to manufacturer recommendations
2. If installed
MAGNETOM
Operator Responsibilities for Maintenance
Reason
Safety of patient and personnel Inspect table movement and
Communication with the patient Function of intercom
Prevent hazards caused by mag-
net
Optimum function within speci-
fications
Safety of patient, personnel, and
system
Optimum function within speci-
fications
Patient protection
M7-000.664.01.04.02
03.12
What is checked:
end switches
Operation of the emergency
stop and safety switches
Operation of the emergency
release
Checking tolerance limits
Marking of the hazard area
Operation of the locking mech-
anism
Inspect the emergency undock-
ing function
Checking the chassis and the
wheels
Function of pneumatic bulb
Operation of the magnet super-
vision and the magnet stop
Function of the gradient super-
vision
Interaction of various system
components according to the
guaranteed system characteris-
tics using the complete QA mea-
surement
Page 8 of 14
CS SD MR
Interval
Annually
Annually
Annually
Annually
Annually
© Siemens, 2009