Performing Matrix Calibration; Calibration Standard; Rs232C Baud Rate - Konica Minolta CA-310 Instruction Manual

Display color analyzer
Table of Contents

Advertisement

3. Performing Matrix Calibration

● Matrix calibration cannot be performed with the memory channel CH00.
(CH00 memory channel is provided for measurement that uses the Konica Minolta's calibration standard.)
● Matrix calibration must be performed for each display type (model).
Characters of displays vary with the display type (model). Because of this, measured values differ even if the
same color is measured. Thus, a different memory channel must be used for each display type (model) to
perform matrix calibration.
● If matrix calibration is performed with a memory channel to which the target color has already been set, that
target color will be deleted.
● If matrix calibration is performed with a memory channel which has already been white-calibrated, the correc-
tion factor of the previous white calibration will be deleted and the correction factor obtained from the matrix
calibration will be set.
● If matrix calibration is performed with a memory channel for which the RGB emission characteristic for the
analyzer mode is to be set, the previous RGB emission characteristic will be deleted and the WRGB set for
matrix calibration will be set as the RGB emission characteristic.
[Operating Procedure]
When the optional 4-Probe Expansion Board CA-B15 is used
Select the probe no. to be white-calibrated. White calibration can be performed independently for each probe
connector ([P1] to [P5]) for each memory channel.
1
Press the
The LCD display section will switch to the menu selec-
tion screen.
2
Press the
key to open the PROBE selec-
tion screen.
Each time the
in the order PROBE → SYNC → ID Name input →
RANGE → Measurement Speed → Number of Digits
→ Calibration Standard→ RS232C Baud Rate →
PROBE.
3
Press the
Each time the
4
Press the
* By default (factory setting), the instrument is set so that [P1] will be selected automatically when the POWER switch is set to ON( | ).
1
2
5∼ 9
5 ∼8
key.
key is pressed, the screen will switch
key to display the probe no. you want to select.
key is pressed, the probe no. switches in the order [P1] ... .
key to confirm the selection.
Menu selection screen
MENU : SELECT
PUSH SPACE KEY
PROBE selection screen
SELECT : PROBE
P1 35881112 A
SELECT : PROBE
P3 35881113
Probe no.
54
5
6
7
8
9
3
Press the
until the desired
probe no. appears.
key

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents