Konica Minolta CA-310 Instruction Manual page 62

Display color analyzer
Table of Contents

Advertisement

1
.
Press the MODE key to select analyzer measurement
mode (RGB).
2
.
Press the MEMORY CH
memory channel where you want to input the RGB emis-
sion characteristic appears.
3
.
Press the
The LCD display section will switch to the analyzer mode RGB emis-
sion characteristic input screen.
4
.
Input the emission characteristic of R.
1 Press the measuring probe against the display, which is now emit-
ting monochrome light of R.
2 Press the
appear on the left of "R".
5
.
Input the emission characteristic of G.
1 Place the measuring probe against the display, which is now emit-
ting monochrome light of G.
2 Press the
key. In the LCD display section, the "*" mark will appear on the left of "G".
6
.
Input the emission characteristic of B.
1 Place the measuring probe against the display, which is now emit-
ting monochrome light of B.
2 Press the
appear on the left of "B".
7
.
Enter the target color (W)
1 Place the measuring probe against the display, which is now emit-
ting the target color(W).
2 Press the
appear on the left of "W".
8
.
Press the
The RGB emission characteristic for the display's analyzer mode and target color (W) will be set.
* Steps 4 to 7 can be performed in any order.
* Pressing the
,
,
or
* To cancel emission characteristic setting, press the
* To change the target color you set, change it as explained in "1. Setting/Changing the Target Color by Measurement" (page 63).
Even if the target color is changed, the currently set RGB emission characteristic for display's analyzer mode will remain unchanged.
* Pressing the MR key displays "100.0" as the target color value for R, B and G.
<Error Messages in LCD Display Section>
● "E1"
• Cause 1 : The display's RGB emission characteristic has never been input for the currently selected
memory channel since shipment from the factory.
Action
: This error will disappear if you enter the emission characteristic.
• Cause 2 : The currently used measuring probe is different from the
one that was used to input the display's RGB emission
characteristic and target color (W).
Action
: Connect the same probe as the one used to input the
display's RGB emission characteristic and target color (W).
and
key.
key. In the LCD display section, the "*" mark will
key. In the LCD display section, the "*" mark will
key. In the LCD display section, the "*" mark will
key.
key before pressing the
key before pressing the
keys until the
key allows you to re-enter the emission characteristic.
key.
... For other error messages, refer to page 103.
60
Memory channel
CH01 EXT
Ad P1A
[
]
CH01
ALZ
P1
W
R
G
B
CH01 EXT
Ad P1A
E1 [
]