Item
Receptor
Measurement area
Acceptance angle
Measurement distance
Luminance
Display range
Chromaticity
Measurement range
Accuracy
(for
white)*1
Luminance
Repeatability(2 )
*1
Measurement range
Accuracy
*1
(temperature:23°±2°,
relative humidity:
(40±10)%))
Chromaticity
Repeatability(2 )
*1
Measurement
xyL
speed*3
V
Digital
Display
Analog
LCD
SYNC mode
Object under measurement
Memory channel
Analyzer function
Interface
Multi-point Measurement
Operating temperature/humidity range
Storage temperature/humidity range
Input voltage range
Main body
Size/weight
Probe
Standard accessories
Optional accessories
*1
The chromaticity and luminance are measured under Konica Minolta's condition (standard LCD(6500K, 9300K) is used).
*2
The luminance for monochrome is measured when the reading of luminance for white is 120 cd/m
*3 Measuring probe connected to probe connector P1 only, used USB(used RS-232C Baud rate: 38400 bps)
*4
"
" represents alternating current available at power outlets.
The specifications and drawings given here are subject to change without prior notice.
LED Universal Measuring Ø10 Probe
Detector: Silicon photo cell
Ø10 mm
±5°
30±5 mm
2
0.0001 to 3000 cd/m
Displayed in 4 or 3-digit value (Can be chosen)
2
0.0150 to 3000 cd/m
2
0.0150 to 0.2999 cd/m
±4%±0.0045 cd/m
2
0.3000 to 29.99 cd/m
±3%±0.0030 cd/m
2
30.00 to 3000 cd/m
±2%±0.0030 cd/m
2
0.0150 to 0.2999 cd/m
1% + 0.0030 cd/m
2
0.3000 to 2.999 cd/m
0.2% + 0.0030 cd/m
3.000 to 3000 cd/m
2
0.1% + 0.0030 cd/m
2
0.1500 to 3000 cd/m
2
0.1500 to 14.99 cd/m
±0.005 for white
2
15.00 to 59.99 cd/m
±0.004 for white
2
60.00 to 3000 cd/m
±0.003 for white
2
120 cd/m
±0.002 for white
(±0.004 for
2
0.1500 to 0.2999 cd/m
0.010
2
0.3000 to 0.5999 cd/m
0.004
2
0.6000 to 1.499 cd/m
0.002
2
1.500 to 3000 cd/m
0.001
2
0.0150 to 0.2999 cd/m
4(3.5) times/sec.
2
0.3000 to 5.999 cd/m
5(4.5) times/sec.
2
6.000 to 3000 cd/m
20(17) times/sec.
xyL
, TΔuvL
, RGB analyze, XYZ, u'v'L
V
V
Δ xΔyΔ L
, R/G B/G ΔG, Δ R B/R G/R
V
16 characters by 2 lines (with backlight)
NTSC, PAL, EXT, UNIV, INT
Vertical synchronization frequency: 40 to 200 Hz
100 channels
Standard function
USB; RS-232C (38,400 bps or below)
Max. 5 points (Use 4-Probe Expansion Board CA-B15)
Temperature: 10 to 28°C; relative humidity 70% or less with no condensation
Chromaticity change ±0.002 for white, ±0.006 for monochrome from reading of Konica Minolta's standard LCD *1, 120 cd/m
with 23°C 40%
0 to 28°C: relative humidity 70% or less with no condensation 28 to 40°C: relative humidity 40% or less with no condensation
100-240V
, 50-60 Hz, 50 VA *4
340(W)×127(H)×216(D) mm / 3.58 kg
Ø49×236 mm / 550 g
AC power cord, PC Software for Color Analyzer CA-SDK, Small Hood CA-HS10
4-Probe Expansion Board CA-B15, Universal Measuring ø10 Probe CA-PSU32(2 m) CA-PSU35(5 m), Small Hood CA-HS10,
Small Lens Cap CA-HS11, USB cable IF-A27
2
2
2
2
2
2
monochrome)*2
V
2
111
Luminance change: ±2% of reading for white ;
2
,