Konica Minolta CA-310 Instruction Manual page 82

Display color analyzer
Table of Contents

Advertisement

<Error Messages in LCD Display Section>
For other error messages, refer to page 101.
● "OVER"
• Measurement is not possible since the instrument's measurement range
is exceeded by the measured value.
In the case of analyzer mode, the instrument's measurement range or
display range (100,000%) is exceeded by the measured value.
● "E1"
• Cause : In the case of xyLv, T∆uvLv, u'v'Lv or XYZ measurement mode,
the currently used measuring probe is different from the one used
to perform user calibration and set the target color. In the case of
analyzer mode, the currently used measuring probe is different
from the one used to input RGB emission characteristic for analyzer mode and set the target
color (W).
• Action: Use the same probe as the one used to input the RGB emission
characteristic and set the target color. Alternatively, input the RGB
emission characteristic and set the target color using the currently
used measuring probe.
● "E2"
• Cause : An error has occurred due to shift of the zero point because the ambient temperature has changed
since zero calibration.
• Action: Perform zero calibration.
* Measurement can still be performed even if "E2" is currently displayed.
* "E2" will not appear if "E1" is currently displayed.
[For Flicker Mode**]
● "FLICKER ERROR UNDER"
• Measurement is not possible since Lv is below 0.1 cd/m
a LED Flicker Measuring ø10 Probe(CA-PS32/35) is coneected.)(white
calibration with Konica Minolta's calibration standard).
● "FLICKER ERROR OVER"
• Measurement is not possible since flicker value is beyond 999.9%.
● "FLICKER ERROR VSYNC OVER"
(EXT is selected as the SYNC mode)
• Measurement is not possible since the frequency of the vertical syn-
chronizing signal input to this instrument is beyond 130 Hz.
(INT is selected as the SYNC mode)
• Measurement is not possible since the currently set vertical scanning
frequency is beyond 130 Hz.
● "FLICKER ERROR PROBE TYPE"
• Cannot measure as the measured probe is not for LCD Flicker Measuring.
**Flicker Mode is a function which can be used only when LED Flicker Measuring ø27 Probe or LED Flicker Measuring ø10 Probe is
connected.
2
2
(0.3 cd/m
when
80
OVER
CH01 NTSC Ad P1A
E1 [
]
CH01 NTSC Ad P1A
E2 [
]
FLICKER ERROR
UNDER
FLICKER ERROR
OVER
FLICKER ERROR
VSYNC OVER