6.3 Clean the housing

Warning!
• Risk of electric shock. Switch the device off completely before cleaning (see
"Switch device off completely" on page 62).
• Ensure that no liquids get into the device. Never immerse the device in
disinfectants or other liquids, otherwise damage to the device and thus a hazard
to users and patients may result.
Proceed as follows to clean the housing.
1. Wipe down the device and the power cord with a soft damp cloth. The therapy device
must be completely dry before the device is started up.
6. Remove the fan filter on the front of the device and clean it as described in
"9.3 Change filter".
7. Put the fan filter back in.

6.4 Clean coarse dust filter/change fine filter

1. Take off the filter compartment lid as described in "9.3 Change filter".
2. Remove the coarse dust filter from out of the filter compartment lid and clean it with
clean running water until there are no residues.
3. Change the fine filter if required.
4. Allow the coarse dust filter to dry. The coarse dust filter must be completely dry before
the device is started up.
5. Put the coarse dust filter back in and close the filter compartment lid.
Note:
The fine filter cannot be cleaned. It is changed every 1000 operating hours.
70
EN
Hygiene treatment
2. Take off the filter compartment lid.
3. Remove the coarse dust filter as described in
"9.3 Change filter".
4. Clean the filter compartment lid under running water
until there are no residues. Then dry it carefully.
5. Insert the coarse dust filter and the filter compartment
lid as described in "9.3 Change filter" on page 97.