Chapter 5 To Trigger the Oscilloscope

TimeOut Trigger (Optional)

In timeout trigger, the instrument triggers when the time interval (△T) from when
the rising edge (or falling edge) of the input signal passes through the trigger level to
when the neighbouring falling edge (or rising edge) passes through the trigger level
is greater than the timeout time set, as shown in the figure below.
Trigger Type:
Press Type, rotate
trigger setting information is displayed at the upper right corner of the screen.
For example,
source is CH1; the trigger level is 0.00 V.
Source Selection:
Press Source to open the signal source list and select CH1-CH4 or D0-D15. For the
details, please refer to the introduction in "Trigger Source". The current trigger
source is displayed at the upper right corner of the screen.
Note:
Edge Type:
Press Slope to select the type of the first edge of the input signal that passes
through the trigger level.
trigger level.
trigger level.
level.
Timeout Time:
Timeout time represents the minimum time that the clock signal must be in the idle
state before the oscilloscope starts searching for the data meets the trigger condition.
Press TimeOut to set the timeout time of timeout trigger and the range is from 16
MSO1000Z/DS1000Z User's Guide
nd Quality Products Online at:
T
Time Out
to select "TimeOut" and press down
Select channel with signal input as trigger source to obtain stable trigger.
: start timing when the rising edge of the input signal passes through the
: start timing when the falling edge of the input signal passes through the
: start timing when any edge of the input signal passes through the trigger
GlobalTestSupply
www.
Figure 5-8 TimeOut Trigger
. The trigger type is timeout trigger; the trigger
Time Out< T
. At this point, the
.com
sales@GlobalTestSupply.co
RIGOL
5-19