RIGOL
Press Delay-Cal and rotate
-100 ns to 100 ns.
Note:
The delay calibration setting is related to the instrument model and the
current horizontal time base. The larger the horizontal time base is, the larger the
setting step will be. Take MSO1104Z as an example; the steps under different
horizontal time bases are as shown in the table below.
Table 7-1 Relationship between the delay calibration step and horizontal time base
Horizontal Time Base
5 ns
10 ns
20 ns
50 ns
100 ns
200 ns
500 ns
1 μs to 10 μs
Note:
When the horizontal time base is equal to or greater than 10 μs, the delay
calibration time cannot be adjusted.
7-6
nd Quality Products Online at:
to set the delay calibration time. The range is from
Delay Calibration Time Step
100 ps
200 ps
400 ps
1 ns
2 ns
4 ns
10 ns
20 ns