SATO CW408 Operator's Manual page 65

Table of Contents

Advertisement

Preliminary
3.16 SERVICE MODE (Cont'd)
1. When SENSOR ADJUST is selected in the SERVICE MODE menu, press the FEED button to switch to
the next setting options as shown above. To return to the previous setting option, press the LINE and
FEED buttons simultaneously.
2. When the desired setting option is displayed, press the LINE button to select an item or to set the value
and then press the FEED button to save the setting.
Menu
SENSOR ADJ
MODE SELECT
SENSOR ADJ
MODE SELECT
AUTOMATIC
MANUAL
SENSOR ADJ(AUTO)
SENSOR SELECT
SENSOR ADJ(AUTO)
ALL
SENSOR SELECT
SENSOR ADJ(AUTO)
I-MARK
SENSOR SELECT
SENSOR ADJ(AUTO)
SENSOR SELECT
GAP
PAPER
SENSOR ADJ(AUTO)
I-MARK SUCCESS
GAP
SUCCESS
PAPER SUCCESS
SENSOR ADJ(AUTO)
TEST PRINT
YES
NO
TEST PRINT
SIZE SELECT
LARGE
SMALL
CW408 Operator Manual
SENSOR ADJUST MODE
To set the sensor adjustment method.
AUTOMATIC or MANUAL can be selected.
The initial setting is AUTOMATIC.
To select the sensor for automatic sensor adjustment.
ALL: Adjust I-mark sensor, Gap sensor and Paper sensor.
I-MARK: Adjust I-mark sensor.
GAP: Adjust Gap sensor.
PAPER: Adjust Paper sensor.
The initial setting is ALL.
This screen shows the result of adjustment for the sensor selected in sensor
adjustment (AUTO) selection screen.
The screen displays SUCCESS when automatic adjustment succeeded.
The screen displays FAILURE when automatic adjustment failed.
When ALL is selected for SENSOR SELECT:
The screen displays the result of adjustment for I-MARK, GAP and PAPER.
When I-MARK is selected for SENSOR SELECT:
The screen displays the result of adjustment for I-MARK.
When GAP is selected for SENSOR SELECT:
The screen displays the result of adjustment for GAP.
When PAPER is selected for SENSOR SELECT:
The screen displays the result of adjustment for PAPER.
Buzzer emits three beeps and goes to TEST PRINT screen.
This screen confirms issuance of a test print for the result of the auto sensor
adjustment.
The initial setting is YES.
To set the test print size of sensor adjustment result.
In this screen, you can choose to print the test results in LARGE or SMALL
print widths.
SMALL: Test print in 50 mm (1.97") wide
LARGE: Test print in 104 mm (4.09") wide
The initial setting is LARGE.