S/Average Menu - Hach POLYMETRON 9125 Operator's Manual

Conductivity/resistivity measurement
Table of Contents

Advertisement

Transmitter 9125 – Conductivity / resistivity measurement
AVERAGE
AVERAGE : 9
TEST :
TEST :
Select
TEST
AVERAGE
28.3 mS/cm
REAL
27.9 mS/cm
Page 52

S/AVERAGE Menu

S/AVERAGE menu program a moving average on the
measurement.
AVERAGE : define the number of measures to
calculate the average.
TEST : visualize the difference between a measure
done with or without average.
Note :
The measurement cycle lasts 4 seconds if there is
a temperature measurement and only 2 seconds
without temperature measurement.

Hide quick links:

Advertisement

Table of Contents
loading