Transmitter 9125 – Conductivity / resistivity measurement
PARAMETERS Menu
PARAMETERS
DATE
: 01/01/01
SLOPE : 100%
COEFF : 0.50
T
: -0.0°C
HISTORIC Menu
HISTORIC
DATE
: 01/01/01
SLOPE : 100%
Page 36
This menu displays the calibration parameters of the
conductivity measurement.
Date of last calibration. This date is the date the user
has entered after a conductivity calibration ( electric, 2
points, 1 point).
The other parameters are those which are displayed
when the date is registered :
Slope of the last calibration which is a correction
factor of the conductivity probe slope.
Si la mesure est effectuée en TDS, le coefficient
de proportionnalité entre la conductivité et la
concentration.
Drift of the temperature measurement.
This
menu
displays
corresponding to the last two conductivity calibrations
and allows to follow the probe clogging.
the
conductivity
slopes