7.

Error messages

ERROR MESSAGE
10.3
mS/cm
Pt100/Pt1000
SHORT CIRCUIT
12.7
mS/cm
Pt100/Pt1000
OPEN CIRCUIT
60
.cm
MEASURE
TOO LOW
15.2
mS/cm
MEASURE
TOO HIGH
15.2
mS/cm
POLARIZATION
TOO HIGH
Note :
In manual acquittal, in order to suppress an error
message press ENTER after correcting the default.
DESCRIPTION/POSSIBLE CAUSE
Sensor not correctly connected
Temperature sensor damaged
Replace it if necessary.
Sensor not correctly connected
Temperature sensor damaged
Replace it if necessary.
The resistivity value is inferior to the lower limit of the
.
measuring range
The conductivity value is superior to the upper limit of
.
the measuring range
The electrode polarization is too high
.