Transmitter 9125 – Conductivity / resistivity measurement
ALARM 3
System
MODE : Syst
ACCEPT : Manu
RELAY : NO
Select
ALARM 4
Timer
MODE : Timer
INTERV : 2440 mn
Impul. Nb : 1
Ton : 005 s
Toff : 005 s
TmA : 00 mn
Select

ALARM 3 (Alarm system)

In case of an alarm 3, choice between
alarm or alarm system function.
MODE :
No
Limit
System
In case of an alarm system, choice between an
automatic accept or a manual accept.
ACCEPT :
Auto
Manu
Choice between relays normally open or closed.
RELAY :
NO
NC

ALARM 4 (Timer)

In case of an alarm 4, choice between a limit or a
timer function.
MODE :
No
Limit
Timer
INTERV : interval between two cleaning cycles in
minutes.
Impul. Nb : number of pulses during a cleaning
cycle.
Ton :
time
when
seconds.
Toff :
time when relay is desactivated, in
secondes.
TmA :
hold time for the analogue outputs in
minutes.
a limit
relay
is
activated,
in