S/Rs485 Menu - Hach POLYMETRON 9125 Operator's Manual

Conductivity/resistivity measurement
Table of Contents

Advertisement

Transmitter 9125 – Conductivity / resistivity measurement
RS485
4
N° : 4
BAUD : 9600
PARIT. : odd
BIT STOP : 1
SWAP WORD : NO

S/RS485 Menu

This option requires the RS485 kit.
Monec number (0...32)
BAUD
300/600/1200/2400/4800/9600/19200
Transmission speed in bauds
PARIT.
- Without parity bit : No
- With odd parity bit : Odd
- With even parity bit : Even
BIT STOP
- 1 bit stop
- 2 bits stop
SWAP WORD Allow
weight », « light weight » size during
the manipulation of the real variable
(float type). Some equipment need
this reverse to read correctly the real
size data.
The communication protocol is MODBUS/JBUS.
The instrument may be equipped with a RS485 board
(optional) (see MODBUS 9100 manual).
to
reverse
the
« strong
Page 49