SYSTEM PROGRAM 2 TCe
SAMPLE IN
8.3.3 Calibration Data
DECIMAL PLACES
TOC/TC CALIBRATION 1
TOC/TC CALIBRATION 2
TOC/TC CALIBRATION 3
TIC CALIBRATION 1
TIC CALIBRATION 2
TIC CALIBRATION 3
8.3.4 Sequence Program
8.3.4.1
Average Program
LOG AVERAGE
AVERAGE UPDATE
SAMPLE IN defines the amount of sample (in pulses) to be
2p , P-V
injected into the reactor in TC mode at range 2.
DECIMAL PLACES setting defines the number of decimal
3
places (0, 1, 2 or 3) system displays in the reaction results and
in the relevant system menus.
The default unit for reaction results is mg/l "milligrams per liter".
Depending on application requirements and preferences, it is
possible to configure the system to display the results in ppm
"parts per million", in µg/l "micrograms per liter" or in ppb "parts
per billion". For further information, contact your local distributor
or the manufacturer.
Calibration menus are used to install the calibration data and
thus to calibrate the BioTector in factory. TOC CALIBRATION 1,
2 and 3 shows the calibration curves for range 1, 2 and 3
respectively. In TC and VOC systems, the TOC CALIBRATION
1/2/3 menus are named as TC CALIBRATION 1/2/3.
The first column shows the number of calibration points, the
second point [%] shows the calibration factors automatically
calculated from the calibration data, the third column
STANDARD shows the standard solution concentrations and the
fourth column shows the un-calibrated result (mgC/l) measured
and calculated from the CO
System calibration curves should not be modified by the user on
site unless a recalibration of the system is required at a different
analysis range.
LOG AVERAGE function enables (YES) or disables (NO by
NO
default) the 24-hour averaging of the reaction results in the
reaction archive.
The reaction results obtained from manual samples are not
included in the averaging calculations.
If the LOG AVERAGE above is enabled (YES), the system
00:00
calculates a 24 hours average of the reaction results and
displays it in the reaction archive for each stream at the time
programmed for AVERAGE UPDATE (00:00, HR:MIN, by
default).
analyzer readings.
2
Page 112