Specifications - ConMed sabre 180 Service Manual

Electrosurgical unit
Table of Contents

Advertisement

•The heat dissipation capability of the heat sink is
severely impaired by activating the Sabre 180 in
other than its normal operating position. There
are no tests requiring operation in any other posi-
tion.

1.2 SPECIFICATIONS

MAINS INPUT REQUIREMENTS:
•Single Phase AC / 48-62 Hz / 200W
•Power Cord: 3 conductor, #18 AWG (1.5mm 2 ) Cu, CEE22 250V 6A Mains Connector
CATALOG NO.
60-5800-001
60-5800-002
60-5800-003
MAINS FREQUENCY LEAKAGE:
•Chassis: <100 uA
•Patient: <10 uA
RF OUTPUT SPECIFICATIONS:
POWER
MODE
WATTS
CUT PURE
CUT BLEND 1
CUT BLEND 2
MONO. COAG
BIPOLAR COAG
RF LEAKAGE: <100 mA, RF ISOLATED
DUTY CYCLE: 15s ON, 30s IDLE
COOLING: Natural convection, no fan
MAINS VOLTAGE
VRMS
RATED
MIN
120
104
100
230
208
MAX.
RATED
CARRIER
LOAD
OHMS
100
500
65
500
50
500
80
500
50
50
F
•Consult the factory for advice before making
any modifications to the unit.
•Ensure that the RUN-RST-CAL switch is set to
RUN and that the two top cover screws are tight-
ened before returning the unit to service.
MAX
127
90
110
254
IEC Class 1, Type CF, Defibrillator Proof
CREST FACTOR
FREQ.
50% POWER
KHz
RATED LOAD
417
1.8
417
2.3
417
3.5
540
7.0
1050
2.4
CURRENT
ARMS
MAX
IDLE
2.4
0.5
2.9
0.6
1.3
0.3
MAX. OPEN
CIRCUIT
PULSE
VOLTAGE
REPETITION
VPP
FREQ. KHz
1600
CONTIN.
1650
20 KHz
1750
20 KHz
6500
33 KHz
400
20 KHz
1-5