1.4 EXPLANATION OF SYMBOLS

CONTROL PANEL
PURE CUT - WAVEFORM WITH MINIMUM
THERMAL DAMAGE AND HEMOSTASIS
BLEND 1 - CUT WAVEFORM WITH
MODERATE HEMOSTASIS
BLEND 2 - CUT WAVEFORM WITH
HEAVY HEMOSTASIS
MONOPOLAR COAGULATION
BIPOLAR COAGULATION
MACHINE FAULT - UNIT HAS
DISABLED ITSELF. REFER TO
ACCOMPANYING DOCUMENTS
RETURN MONITOR ALARM -
MONOPOLAR OUTPUT IS DISABLED
SINGLE FOIL - USED WITH ELECTRODES
THAT DO NOT MONITOR CONTACT QUALITY
DUAL FOIL - SETS MONITOR TO USE
RESISTANCE - IN DUAL FOIL MODE,
INDICATES RELATIVE RESISTANCE
OF RETURN ELECTRODE CONTACT
WITH THE PATIENT
TOP COVER
IPX1
DRIP PROOF - ENCLOSURE PROTECTED
AGAINST VERTICALLY FALLING WATER
DROPS WHEN EQUIPMENT IS IN A
POSITION OF NORMAL USE
POWER SWITCH
I
POWER ON - POWER SWITCH
O
POSITION TO TURN EQUIPMENT ON
POWER OFF - POWER SWITCH
POSITION TO TURN EQUIPMENT OFF
THIS EQUIPMENT INTENTIONALLY SUPPLIES
NON-IONIZING RF ENERGY FOR
PHYSIOLOGICAL EFFECT.
1-10
OUTPUT PANEL
HANDSWITCHED OUTPUT - CONNECTION
FOR HANDSWITCHED MONOPOLAR
ACCESSORIES
FOOTSWITCHED OUTPUT - CONNECTION
FOR FOOTSWITCHED MONOPOLAR
ACCESSORIES
HIGH VOLTAGE CAUTION -
EQUIPMENT HAS HIGH VOLTAGE OUTPUT
CAPABLE OF CAUSING PHYSIOLOGICAL
EFFECTS
BIPOLAR OUTPUT - CONNECTION FOR
BIPOLAR ACCESSORIES
FOOTSWITCHED BIPOLAR OUTPUT
TYPE CF - PATIENT CONNECTIONS ARE
ISOLATED FROM EARTH AND RESIST THE
EFFECTS OF DEFIBRILLATOR DISCHARGE
RETURN ELECTRODE - CONNECTION FOR
MONOPOLAR DISPERSIVE ELECTRODE
RF ISOLATED - PATIENT
F
CONNECTIONS ARE ISOLATED FROM
EARTH AT HIGH FREQUENCY
CONSULT ACCOMPANYING DOCUMENTS
PRIOR TO PLACING EQUIPMENT
IN SERVICE
REAR PANEL
FOOTSWITCH CONNECTOR
ACTIVATION TONE VOLUME
A4 ASSEMBLY
CAUTION - THIS CIRCUITRY
SUPPORTS PEAK VOLTAGES GREATER
THAN 1000V
PROTECTIVE EARTH (GROUND)