Iec 60601-1-2 Table 206; Specifications; Mains Overcurrent Protection; Mains Frequency Leakage - ConMed SYSTEM 5000 Operator's Manual

Electrosurgical unit
Hide thumbs Also See for SYSTEM 5000:
Table of Contents

Advertisement

1.1.5.4 EN/IEC 60601-1-2 Table 206
Recommended separation distances between portable and mobile RF communications equipment
The System 5000™ Electrosurgical Unit is intended for use in an electromagnetic environment in which radiated RF
disturbances are controlled. The customer or the user of the System 5000™ Electrosurgical Unit can help prevent elec-
tromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equip-
ment (transmitters) and the System 5000™ Electrosurgical Unit as recommended below, according to the maximum
output power of the communications equipment.
Rated maximum output
power of transmitter
(W)
0.01
0.1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters
(m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output
rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.
1.2

Specifications

Input Power: 600 watts maximum
Mains Voltage VRMS
Frequency
Hz
Nominal
50-60
115
50-60
100
50-60
230
*Fuses are Type T, High Breaking
1.2.1

Mains Overcurrent Protection

Two fuses for each ESU with ratings as noted in the table above.
1.2.2

Mains Frequency Leakage

Designed to comply with IEC60601-1: 1995.
Patient connections to Neutral: Less than 10 µA.
Chassis to Neutral: Less than 25 µA (120V), Less than 45 µA (240V).
and the System 5000™ Electrosurgical Unit
Separation distance according to frequency of transmitter
150 kHz to 80 MHz
d = 1.2√P
0.12
0.38
1.2
3.8
12
Minimum
Maximum Maximum
104
127
90
110
198
253
(m)
80 MHz to 800 MHz
d = 1.2√P
0.12
0.38
1.2
3.8
12
Mains Current ARMS
5.3
6.2
2.7
800 MHz to 2.5 GHz
d = 2.3√P
0.23
0.73
2.3
7.3
23
Fuses*
T6.3A
T8.0A
T3.15A
1-9

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents