Specifications - ConMed 1000 SES Service Manual

Smoke evacuation system
Table of Contents

Advertisement

2. Attach unit power cord to the receptacle on the
rear panel of the machine. Plug the three prong
power connector into an appropriate power out-
let.
3. Optional: Attach either Remote Activator
Switch or Footswitch by inserting the male
footswitch plug into REMOTE SWITCH recep-
tacle on rear of machine.
4. Attach the CONMED Smoke Evacuation
Pencil or tubing set to the inlet of the CONMED
1000SES
TM
Filter. Ensure that the evacuation
tubing is fully seated on the inlet of the filter.
5. Turn on unit main power by placing the power
switch on the rear panel of the machine in the
"ON" position.
6. Activate the power unit by:
a. Pressing the ON/OFF switch on the
membrane panel,
b. Depressing and releasing the footswitch (if
connected), or
c. Activating either CUT or COAG on the
CONMED Smoke Evacuation Pencil (if
Remote Activator is connected).
7. Adjust the suction level to the desired setting
by pressing the UP and DOWN arrow keys while
the unit is activated. Noise created by the smoke
evacuation power unit may be minimized by
selecting the lowest vacuum setting that effective-
ly clears the sterile field of surgical smoke. When
two indicators are illuminated on the suction con-
trol scale, the suction control setting is the value
between the two illuminated percentages.
8. Deactivate the unit by:
a. Pressing the ON/OFF switch on the
membrane panel,
b. Depressing and releasing the footswitch (if
connected), or
c. Releasing either the CUT or COAG button
on the CONMED Smoke Evacuation Pencil
(if Remote Activator is connected).
9. Replace the CONMED 1000SES
the Filter Life Remaining Scale flashes RED (0%
Life Remaining). Failure to change the filter will
affect the performance of the system.

2.4 Specifications

Flow Rate:
Motor Rating: 115 CFM (3257 LPM)
Static Suction: 97" (246cm) H
Power:
Dimensions:
Weight:
Tubing Sizes:
Filter:
Filtration Efficiency: 99.99995%
Filtration Rate: 0.012 µm
Max. Filter Service Life: 8 Hours
Regulatory Compliance
Units conforming to UL2601-1:
60-6850-001
60-6850-002
60-6850-003
60-6850-004
UL 2601-1 Classifications:
Class 1
Type CF
Ordinary Equipment
Continuous Operation
TM
Filter when
70 CFM (1982 LPM) with
1-1/4" (32mm) hose
48 CFM (1359 LPM) with
7/8" (22mm) hose
O
2
15 AMPS @ 100V/120V
8 AMPS @ 220/240V
Height - 14.25" (36.2cm)
Width - 12.38" (34cm)
Length - 18.25" (46.4cm)
25.5 lbs (11.6 kg)
7/8" (22mm), 1-1/4" (32mm),
and 1/4" (6.4mm)
CONMED 1000SES
TM
VLSI grade ULPA with carbon
120 V 60 Hz
220 V 50 Hz
240 V 50 Hz
100 V 50/60 Hz
2-3