Specifications - ConMed hyfrecator 2000 Operator's Manual

Hide thumbs Also See for hyfrecator 2000:
Table of Contents

Advertisement

PARAMETER
Primary Power Requirements
Model 7-900-100
Model 7-900-115
Model 7-900-220
Model 7-900-230
Dimension and Weight
Depth:
Width:
Height:
Weight:
Functional Characteristics
Mode
Max Power
Max Voltage
Setting (Watts)
HI
35
LO
20
BI
35
Power Accuracy at Rated Load
LO and BI:
0 to 10 W: Greater of 90 mW or 10% of setting
10 to maximum setting: 10% of setting
HI:
Greater of 1 W or 10% of setting
NOTE: Instrumentation errors can be great enough to exceed these limits. For best results, the same instruments
should be used for calibration, as well as checking.
RF Frequency:
Output Wave Shape:
Line Regulation:
Duty Cycle:
Audio:
Display Settings:
Low Output Terminal
0-10 range:
10-20 range:
Bipolar & High Output Terminals
0-35 range:
60 Hz Leakage:
Mains Power Cord:
Environmental
Operating Temperature Range:
Relative Humidity:
Storage Restrictions:
Approvals:
Protection:
CHARACTERISTICS
100V± 10% 50/60 Hz 0.9 A
115V± 10% 50/60 Hz 0.8 A
220V± 10% 50/60 Hz 0.5 A
230-240V± 10% 50/60 Hz 0.5 A
4 inches (102 mm)
8.75 inches (222 mm)
7.5 inches (190 mm)
6 pounds (2.7 kg)
Rated Load
Repetition Frequency
(Vpp)
(Ohms)
8000
1000
3000
500
3000
500
450 KHz ±50KHz (open circuit)
Clipped, damped sinusoidal
± 5% at at max. power for a +/- 10% line voltage
Intermittent: 30 sec. on, 30 sec. off
Adjustable from >45 dba to >65 dba at 1 meter
0.1 watt increments
1 watt increments
1 watt increments
50 microamps max. third wire of line cord open
18 AWG/3 105
10 ft. (3m)
+10 degrees to +40 degrees C
10% to 95% non-condensing
-40 degrees to +70 degrees C, 95% humidity N.C.
UL 544, Medical and Dental Equipment, E68077
EN 60601-1 / IEC 60601-1
BF Defibrillator, Class 1
Section 15 - Specifications
(KHz)
24.4
32.3
32.3
o
C 300V / IEC-320-C13 connector,
Specifications subject to change without notice.
2000
21

Hide quick links:

Advertisement