2.4 CONTROLS, DISPLAYS AND
CONNECTORS
(1) CUT ACTIVATION INDICATOR - Lit
when a cutting output is present.
(2) CUT POWER SETTING - Displays power
setting of selected cut mode.
(3) COAG ACTIVATION INDICATOR - Lit
when monopolar or bipolar coag output is
present.
(4) COAG POWER SETTING - Displays
power setting of selected coag mode.
(5) MACHINE INDICATOR - Lit when an
internal fault has disabled unit. Accompanied by
diagnostic code in power setting windows (2),
(4).
(6) RETURN MONITOR INDICATOR - Lit
when monopolar output is disabled due to a fault
in the return electrode circuit.
(7) RESISTANCE BAR GRAPH- Indicates rela-
tive resistance at dual foil monitoring return
electrode.
1
CUT
WATTS
PURE
BLEND 1
BLEND 2
MODE
13
2-4
2
COAG
MONOPOLAR
MODE
12
FIGURE 2.1 CONTROL PANEL
(8) MONITOR SET PUSHBUTTON - In dual
foil mode, stores present monitor resistance and
clears return monitor alarm (6).
(9) PAD SELECTION - Selects whether return
monitor is to operate with SINGLE foil (non-
monitoring) or DUAL foil (contact quality moni-
toring) return electrodes.
(10) COAG POWER CONTROL - Adjusts
coag power setting (4) of selected coag mode
(11). Does not affect the non-selected mode.
(11) COAG MODE SELECTION
•MONOPOLAR - Coag hand or foot switch
activates coag output at appropriate monopolar
active output.
•BIPOLAR - Coag foot switch activates bipolar
output. Cut hand or foot switch will operate
as in monopolar mode.
(12) CUT POWER CONTROL - Adjusts cut
power setting (2) of selected cut mode (13).
(13) CUT MODE SELECTION - Selects degree
of hemostasis of monopolar cut waveform as
indicated by associated indicator lamps.(1)
3
4
WATTS
SINGLE
BIPOLAR
11
10
5 6
7
MACHINE
RETURN MONITOR
DUAL
MONITOR
PAD
SET
9
8