CONMED 1000SES Smoke Evacuation System against any defects in material or workmanship and will repair or replace (at CONMED’s option) the same without charge, provided that routine maintenance as specified in this manual has been performed using replacement parts approved by CONMED. This warranty is void if the product is used in a manner or for purposes other than intended.
Cord for connection to a standard outlet. Penetration Air) filter whose high-tech patented design captures particulates and micro-organisms A list of available accessories for the CONMED down to 0.012 µm at an efficiency of 1000SES is presented in Section 4.3. Please 99.99995%.
• Using any other filter or accessory not sup- hose assembly. Failure to install a fluid collec- plied by CONMED may cause damage to the tion device could cause filter blockage and system and will void the warranty.
Operating Instructions Section 2.0 2.1 System Controls • FOOTSWITCH / REMOTE SWITCH ACTIVATOR (Membrane Control The electronic system controls on the CONMED Panel) See Figure 1. 1000SES are easy to understand and simple to use. The membrane control panel contains the...
Note: Before installing or removing any filter, CAUTION: Using any other filter or accesso- be sure that the system is turned off and the ry not supplied by CONMED may cause dam- electrical plug has been disconnected. age to the system and may void the warranty.
Motor Rating: 115 CFM (3257 LPM) tacle on rear of machine. Static Suction: 97" (246cm) H 4. Attach the CONMED Smoke Evacuation Pencil or tubing set to the inlet of the CONMED Power: 15 AMPS @ 100V/120V 1000SES Filter. Ensure that the evacuation 8 AMPS @ 220/240V tubing is fully seated on the inlet of the filter.
Remote Power Remote Power Switch Switch Switch Switch Jack Jack Circuit Circuit Power Breakers Breakers Power Cord Cord Connector Connector Cautions Cautions IPX1 Product Product Nameplate Nameplate Figure 2 - Back View Handle Handle Control Control Panel Panel Filter Filter Inlet Inlet Cooling...
The CONMED 1000SES should be visually This section contains information for ordinary inspected at least every year. This inspection maintenance of the CONMED 1000SES Smoke Evacuation System. While the unit has should include checks for: been designed and manufactured to high industry •Damage to the power cord.
2. Replace filter. 3. Vacuum hose clogged. 3. Check or replace Smoke Evacuation Tubing. 4. Motor/Blower obstructed. 4. Call BioMed or CONMED Tech Services at the numbers listed on the inside front cover of this manual. 1. Not plugged into electrical 1.
CON- filter cartridge is used to simplify the installation MED 1000 SES through the exit port on the bot- and removal during filter changes. This filter is tom of the unit via airways in the unit’s insula- completely enclosed to further protect the health- tion.
97" H O. Replacement brushes for this AC motor are available from CONMED The motor/blower is a specially designed high Technical Services (See Section 4.0). Refer to the suction/high flow rate centrifugal air pump. This...
If filter life has been expended, replace the filter. This section provides help in troubleshooting problems if the CONMED 1000 SES is malfunc- Weekly Inspection tioning. Not all failures can be covered in a guide such as this, so the service technician must under- Check the “Service”...
Page 20
Check the mains voltage against the nameplate on the back of the unit. Operating the Problem Potential Cause Corrective Action 1. CONMED 1000 SES 1. Filter not seated completely. 1. Reinstall filter, press firmly into is operational, but suction is place.
Allen wrench. WARNING: Before performing any work on Using a wrench, remove the bolt from the motor the CONMED 1000 SES be sure the system mount bracket. is turned off and the electrical plug has been disconnected from the wall outlet.
6.6.3 Assembly Instructions with new brushes, and reinstall the two (2) Phillips head screws. Tighten down securely. For assembly of the CONMED 1000 SES Smoke Evacuation System, please use the above instruc- Reassemble unit by following the disassembly tions in reverse.