Specifications - ConMed ExcaliburPLUS PC Service Manual

Electrosurgical unit
Table of Contents

Advertisement

Exec
1.2 SPECIFICATIONS
INPUT POWER: 700 W max.
MI
rr
Mains Voltage
Max Mains Current
Mains Circuit
VRMS
ARMS
Breaker Rating
Catalog
Freq.
Nom.
Min.
Max.
Nominal
Max.
Number $
Hz
Mains
Mains Freq.
(+1 Hz)
Idle
160-6290-120
60
120
104
127
0.8
7.0
10A/120V
60-6290-226
60
220
198
242
0.5
3.8
5A/240V
60-6290-230
50
220-230
198
253
0.5
3.8
5A/240V
|60-6290-240
50
240
216
264
0.4
3.5
5A/240V
60-6290-100
50-60
100
90
110
10
84
10A/120V
MAINS OVERCURRENT PROTECTION:
Combined 2-pole circuit breaker/power switch
MAINS FREQUENCY LEAKAGE:
(Per IEC 601-1)
Patient Connections to Neutral:
Less than 10 yA
Chassis to Neutral: Less than 25 wA (120 V)
Less than 45 A
(240 V)
TEC Class I, Type CE, Defibrillator Proof
4 v
VW
REGULATORY CERTIFICATION
Non-Ionizing Radiation
UL Listed, Medical and Dental Eguipment
(E68077)
ISO 9002:1994 Registered Facility
C € 0123
POWER DISPLAY ACCURACY: The greater of
+/- 10% of display or +/-3 watts at rated load.
LINE REGULATION: Power change < 1%/V
for VLINE = Min to Max Voltage
Duty Cycle: 15s on, 30s idle
ENVIRONMENTAL
Operating Conditions: 10°C to 30°C,
95% RH Non-Condensing maximum
MOUNTING RESTRICTION: 2 inch (5cm)
clearance required on each side and above unit for
cooling.
COOLING:
Natural convection, conduction and
radiation. No fan.
Storage Conditions: -40°C to 70°C, altitude -60
to +4500 m MSL when sealed in original poly-
bag, packing material and shipping carton.
Prior to shipment or storage, the unit should be
enclosed and sealed in a polyethylene bag and
placed in original carton using original packaging
materials.
a €
ή

Advertisement

Table of Contents
loading