ConMed Linvatec Advantage Turbo D9920 Manual

ConMed Linvatec Advantage Turbo D9920 Manual

Shaver handpiece

Advertisement

W41-261-004AB.book Page 1 Thursday, September 18, 2014 9:28 AM
ConMed Linvatec Shaver Handpieces
D9920
D9924

Advertisement

Table of Contents
loading

Summary of Contents for ConMed Linvatec Advantage Turbo D9920

  • Page 1 W41-261-004AB.book Page 1 Thursday, September 18, 2014 9:28 AM ConMed Linvatec Shaver Handpieces D9920 D9924...
  • Page 2 ConMed Linvatec. ConMed Linvatec reserves the right to revise this publication and to make changes from time to time in the contents hereof without obligation to notify any person of such revision or changes, unless otherwise required by law.
  • Page 3: Table Of Contents

    W41-261-004AB.book Page i Thursday, September 18, 2014 9:28 AM Table of Contents Page INTRODUCTION Intended Use ......... . 1 Operating Principle .
  • Page 4 W41-261-004AB.book Page ii Thursday, September 18, 2014 9:28 AM Table of Contents Page Handpiece Environmental Requirements ....15 CUSTOMER SERVICE Assistance ......... . . 16...
  • Page 5: Introduction

    INTRODUCTION These shaver handpieces come in two styles; no-button and 2-button functionality. They work with a variety of ConMed Linvatec shaver blades and burs and all are used in conjunction with several drive consoles; the Advantage® System (D3000 Series), E9000® System (E9000 Series), and the PowerPro® System (PRO2000 Series) (all with software version 7.0 or higher).
  • Page 6: Warnings And Precautions

    W41-261-004AB.book Page 2 Thursday, September 18, 2014 9:28 AM Warnings and Precautions DO NOT bypass this section. It contains warnings and precautions that must be thoroughly understood before operating any of the equipment. Lack of understanding or adherence to these warnings and precautions may result in injury or even death to the patient.
  • Page 7: Warnings

    W41-261-004AB.book Page 3 Thursday, September 18, 2014 9:28 AM 1.4.1 Warnings 1. Eye protection is recommended when operating equipment. Eye injury may result. 2. It is the surgeon’s responsibility to be familiar with the appropriate surgical techniques prior to use of the shaver handpieces and associated blades and burs. 3.
  • Page 8: Precautions

    There are no user-serviceable parts inside. 3. Use only associated ConMed Linvatec attachments and accessories. 4. Handle all equipment carefully. If any equipment is dropped or damaged in any way, return it immediately for service.
  • Page 9 (wear), motor seizure, and overheating. 13. If shaver blade clogs, use ConMed Linvatec Declogger (E9315) to declog shaver blade. Shaver blades or burs may also be declogged by removing the inner assembly of the shaver blade from the outer assembly and clearing any tissue from the window or aspiration hole.
  • Page 10: Symbol Definitions

    W41-261-004AB.book Page 6 Thursday, September 18, 2014 9:28 AM Symbol Definitions 1.5.1 Handpiece Symbols Handpiece direction selection button (forward, reverse, oscillate). Will also convert to Speed Increase button (see step 4, “Handpiece Descriptions” on page 8 for more information). Handpiece ON/OFF button. Will also convert to Speed Decrease button (see step 4, “Handpiece Descriptions”...
  • Page 11: Warnings And Information Symbols

    Consult instructions for use Non sterile Date of Manufacture No user service recommended. Refer servicing to qualified ConMed Linvatec service personnel Do Not Resterilize Sterilized using Irradiation STERILIZE Do Not Re-use Temperature Limition Do Not Use if package is...
  • Page 12: Handpiece Descriptions

    W41-261-004AB.book Page 8 Thursday, September 18, 2014 9:28 AM Handpiece Descriptions D9920: No Button Shaver Shown D9924: 2- Button Shaver Shown The information below pertains to the Turbo (D9920, D9924) shaver handpieces. All these handpieces are equipped with sensors that automatically select the proper speed range for the attached blade or bur.
  • Page 13: Preoperative Functional Test

    W41-261-004AB.book Page 9 Thursday, September 18, 2014 9:28 AM Speed button. This enables the surgeon to change the handpiece operating speed from the sterile field. If no button is pressed for 1.5 seconds, the console beeps twice and returns to normal button functionality. 5.
  • Page 14: Cleaning And Sterilizing

    W41-261-004AB.book Page 10 Thursday, September 18, 2014 9:28 AM CLEANING AND STERILIZING Cleaning and Washer/Sanitizer Warnings and Precautions 1. Follow universal precautions for protective apparel when handling and cleaning contaminated instruments. 2. Clean instruments within 30 minutes after use to minimize the potential of blood and debris drying prior to cleaning.
  • Page 15: Handpiece Cleaning Instructions

    W41-261-004AB.book Page 11 Thursday, September 18, 2014 9:28 AM Handpiece Cleaning Instructions Manual Cleaning Clean handpieces as soon as possible after use. 1. Disconnect the handpiece cord from the console and remove all accessories from the handpiece before cleaning. 2. Thoroughly scrub the handpiece and handpiece cord with a clean, soft brush dampened, not saturated, with a mild, pH-balanced detergent.
  • Page 16: Warnings, Precautions And Notes

    W41-261-004AB.book Page 12 Thursday, September 18, 2014 9:28 AM 3.2.1 Warnings, Precautions and Notes 1. The use of disinfecting solutions for an exterior instrument wipe will not sterilize the equipment and is not recommended. 2. Do not “Peel Pack” handpieces for sterilization. Sterilization in a sealed pouch traps moisture which can cause damage.
  • Page 17: Technical Specifications

    W41-261-004AB.book Page 13 Thursday, September 18, 2014 9:28 AM TECHNICAL SPECIFICATIONS I.E.C. Equipment Classification: Class I (intended for connection to Class I surgical console), IPX0 (ordinary) Type of applied part is dependent on surgical console. The system was tested and complies with UL 60601-1, 1st edition, CAN/CSA C22.2 NO.
  • Page 18: Shaver Handpiece Technical Specifications

    W41-261-004AB.book Page 14 Thursday, September 18, 2014 9:28 AM Shaver Handpiece Technical Specifications Advantage Turbo No-button (D9920) and 2-Button (D9924) Shaver Handpieces Operating Speed:* 500 - 12,000 rpm, Forward/Reverse 500 - 4,000 cpm, Single-and Multi-turn Oscillate Torque (Nominal): 35 in. oz. (24.7 N cm) Length: 6.1 in (15.5 cm) Diameter:...
  • Page 19 W41-261-004AB.book Page 15 Thursday, September 18, 2014 9:28 AM Handpiece Environmental Requirements Environmental Operating Storage and Transport Conditions Temperature: Relative Humidity: Non-Condensing Non-Condensing Atmospheric Pressure:...
  • Page 20 Returning products for any reason requires an authorized Service Request (S.R.) number prominently displayed on the box and included on all paperwork. Refer to this number if making inquiries about the repair status. Please call ConMed Linvatec Customer Service and provide the following information to obtain an S.R.
  • Page 21 W41-261-004AB.book Page 1 Thursday, September 18, 2014 9:28 AM...
  • Page 22 W41-261-004AB.book Page 2 Thursday, September 18, 2014 9:28 AM 525 French Road Utica, NY 13502-5994 USA (727) 392-6464 Customer Service: 1-800 237-0169 FAX: (727) 399-5256 International FAX: +1 (727) 397-4540 email: customer_service@conmed.com www.conmed.com All rights reserved. W41-261-004 Rev AB 09/2014...