Page 1
ConMed Linvatec 2330 Millrace Court, Unit 5 Mississauga, ON L5N 1W2 For technical support & loaner requests please call: 1-800-387-9636...
Page 2
TABLE OF CONTENTS LS7700 SERVICE DISCLAIMER ..................3 LS7700 BLOCK DIAGRAM ..................... 4 BILL OF MATERIALS ...................... 5 LS7700 FINAL CHECKOUT PROCEDURES (FCO) ............7 LS7700 USER MANUAL ....................11...
Page 3
In no case shall ConMed Linvatec be liable for any direct, indirect, special, incidental or consequential damages, based upon any theory of liability arising from a product serviced in any way by other than ConMed Linvatec certified technicians.
Page 4
ConMed Linvatec Endoscopy 7416 Hollister Avenue Goleta, CA 93117 -2583 (805) 571-8100 CONFIDENTIAL This document contains propietary information and shall not be copied, reproduced, disclosed to others or used in any way without the prior written consent of ConMed Linvatec.
Page 12
ConMed Linvatec. ConMed Linvatec reserves the right to revise this publication and to make changes from time to time in the contents hereof without obligation to notify any person of such revision or changes, unless otherwise required by law.
Do not use the light source with cord. Always inspect cords for signs of incompatible equipment or accessories that excessive wear or damage. If wear or are not authorized by ConMed Linvatec. damage are found, discontinue use and Doing so may void certifications and/or replace immediately.
Ensure all items listed below are included in the box and are free from damage. If any damage is noted contact your ConMed Linvatec Customer Service. Save ALL packaging materials; they may be needed to verify any claims of damage by the shipper.
flammable Light Guide OPEN anesthetics Type BF equipment Light Guide LOCK No user service recommended. WEEE (Waste Electronics and Refer servicing to qualified Electrical Equipment) Symbol. ConMed Linvatec service Regarding European Union personnel end-of-life of product...
❸ ❹ ❺ ❷ ❻ ❶ ❹ “BRIGHTNESS” Buttons — System Features Allows the user to increase or decrease the light output. ❺ Light Guide Lever — Allows the 1.5.1 Front Panel user to open the LGR for insertion / removal of a fiber optic light guide.
® Surgeon’s Assistant (SA) to enable external control by an NA or SA. ❹ “CAMERA” Connector — This connector accepts an ethernet cable only from an optional ConMed Linvatec IM4000 series camera so that camera can control the light source.
SYSTEM INSTALLATION Operation of Light Source and OPERATION Use the following sequence to operate the light source: Installation Instructions 1. Press the POWER button to turn the light source on. The power button 1. Attach the hospital grade power cord to the LED will illuminate green and the power cord receptacle on the back panel of OLED displays “STAND-BY”...
Warning: Before taking the light source out Description of Features of stand-by mode, ensure that the output end of the light guide or endoscope is not in proximity to surgical drapes, flammable Light Source OFF: objects, or to the patient’s skin. Operating •...
Page 23
Lamp Replacement” on page 10 . • To enable Auto Shutter control, the ethernet cable must be connected • Replace only with ConMed Linvatec between an IM4000 series ConMed Lamps. Use of unauthorized lamps will Linvatec camera and the light source. void certifications and/or warranties and may affect safety and performance.
MAINTENANCE Lamp Replacement Regular maintenance of your light source is The following maintenance procedures should limited to: 1) keeping the intake vents free of only be performed by qualified service dust and dirt and 2) replacement of the lamp, personnel. Read all warnings and cautionary when needed.
Page 25
1. Depress the POWER button to the off position. 2. Unplug the light source. Allow to cool before proceeding. 3. Flip the latch on the lamp access door towards the back of the unit and pull the door open. 6. After installation is complete, close and secure the door.
Fuse Replacement 1. Depress the “POWER” button to the off position and unplug the light source. 2. Using a small screwdriver or similar tool, gently release the retention tab on the fuse drawer and slide it out. 4. Return fuse drawer to compartment, making sure alignment tab is on top.
♦ Turn power off. Ensure intake expired. vents are not blocked. Allow system to cool for five minutes. Turn power back on. If lamp does not ignite, replace lamp assembly. ♦ Use of unauthorized lamp module. Use only ConMed authorized replacement lamp modules.
Page 28
Symptom Possible Cause Corrective Action ♦ Brightness is set to minimum or ♦ Increase lamp brightness. Low light output. too low of an output position. ♦ Fiber optic light guide surface(s) ♦ Replace light guide. is contaminated or damaged. ♦ Lamp life exceeds warranted ♦...
The power supply is a switch mode type with a Theory of Operation built-in, high voltage lamp ignitor and current regulator. The fans are powered from an The xenon light source uses a replaceable short auxiliary 12VDC power output of the main arc xenon lamp with an integral elliptical supply.
TECHNICAL INFORMATION Light Source I.E.C. Equipment Classification: Type BF, Class 1, continuous operation Typical Operating Requirements: Input Voltage: 100 - 240 VAC Input Line Frequency: 50/60Hz Power Consumption: 500 VA max Fuse: Fast Blow 6.3 Amps, 250 V Intensity Control: Manual or Automatic Weight: 16.0 lbs.
Page 31
Power Cord Requirements: 100/120 Volt (Cat. No. C7104) Use only a listed (UL, CSA) detachable power cord manufactured to the following specifications. • Plug End • NEMA 5-15P hospital grade, 15 amps, 125V • Receptacle End • IEC 320/CEE-22, 6 amps, 250V/15 amps, 125V •...
Page 32
Table 1: Guidance and Manufacturer’s Declaration - Electromagnetic Emissions The ConMed Linvatec LS7700 Light Source is intended for use in the electromagnetic environment specified below. The customer or the user of the LS7700 Light Source should assure it is used in such an environment.
Page 33
Table 2: Guidance and Manufacturer’s Declaration - Electromagnetic Immunity The ConMed Linvatec LS7700 Light Source is intended for use in the electromagnetic environment specified below. The customer or the user of this LS7700 Light Source should assure that it is used in such an environment.
Page 34
Table 2: Guidance and Manufacturer’s Declaration - Electromagnetic Immunity The ConMed Linvatec LS7700 Light Source is intended for use in the electromagnetic environment specified below. The customer or the user of this LS7700 Light Source should assure that it is used in such an environment.
Page 35
Table 2: Guidance and Manufacturer’s Declaration - Electromagnetic Immunity The ConMed Linvatec LS7700 Light Source is intended for use in the electromagnetic environment specified below. The customer or the user of this LS7700 Light Source should assure that it is used in such an environment.
Page 36
Table 3: Recommended Separation Distances Between Portable and Mobile RF Communications Equipment and the LS7700 Light Source The ConMed Linvatec is intended for use in an electromagnetic environment LS7700 Light Source in which radiated RF disturbances are controlled. The customer or the user of the...
R.G. number prior to returning any product for repair, contact ConMed Linvatec Customer repair: Service at 800-925-4255 or your ConMed Linvatec Sales Representative. Outside the U.S. • Product Number contact your local ConMed Linvatec Representative.
Page 42
Linvatec Europe B1070-Brussels, Belgium All rights reserved. Printed in USA W41-129-004 Rev. AA 06/2007...