2.6 Contact output signal
Contact output signal 1
Contact output signal 2
Contact output signal 3
Product condition
(Set condition)
Digital output signal
(N.O type)
Digital output signal
(N.C type)
HRS Series
There are three contact output signals. As the default setting, contact output
signal 1 is for operating condition (N.O type), contact output signal 2 is for
remote signal (N.O type), and contact output signal 3 is for alarm signal
(N.C type). Refer to Table 2-6-1. .Depending on the product condition,
contact output signal is turned on (closed) or turned off (open).
The signals can be customized depending on the customer's application.
The Table 2.6-2 shows operation of contact output signal.
【Tips】
All contact output signals are turned off (open) when the power is not supplied.
Table2.6-1 Contact output signal (Default setting)
Class of signal
Description
Display
Run status signal
Remote signal
Alarm signal
Chapter 2
Contact input/output communication
Signal configuration
Description
N.O type
RUN
RMT
N.O type
ALM
N.C type
HRX-OM-M091
Remarks
Display
A
A
B
Generation(set)
Non generation(not set)
ON
OFF
ON
OFF
2.6 Contact output signal
2-19