SMC Networks HRS012-A Operation Manual page 79

Thermo chiller
Hide thumbs Also See for HRS012-A:
Table of Contents

Advertisement

6.3.2
Setting and checking
1.
Press and hold the [MENU] key for 2 sec.
Repeat pressing the key until the setting screen for alarm buzzer sound [
on the digital display.
PV
SV
Changing of communication error Setting and checking
2.
Display the screen of the communication error by pressing the [SEL] key 10 times.
The set screen of "changing of communication error" is displayed on the digital display.
PV
SV
3.
Select changing of operation when communication error from the table below with [▲] key
or [▼] key, and confirm by pressing "SEL".
Set values
The monitoring time of communication error Setting and checking
4.
Press the [SEL] key once.
The set screen of the "monitoring time of communication error" is displayed on the digital
display.
PV
SV
HRS Series
[MENU]
Press
and hold
PV
SV
Table 6.3-2 List of set values
Explanation
This alarm signal is not detected.
Operation continues when this communication
error is generated.
Operation
is
stopped
communication error is generated.
Chapter 6
[MENU]
Press
and hold
PV
SV
when
this
HRX-OM-M091
Communication alarm function
] appears
[MENU]
Press
and hold
PV
SV
Initial value
(Default setting)
6.3 Setting and checking
6-3