Sec; Configuring The Meter; Selecting The Input Type - Omega DP25B-S User Manual

Strain gage panel meter
Hide thumbs Also See for DP25B-S:
Table of Contents

Advertisement

SECTION 4. CONFIGURING THE METER
Note
Refer to Table 6-1 for a summary list of menu configuration.
For first-time users: Refer to the QuickStart Manual for basic operation and
set-up instructions.

4.1 SELECTING THE INPUT TYPE

To select your appropriate input type signal, follow these steps:
Note
Before proceeding, set the input DIP switch settings at the back of your
meter. (Refer to Table 2-3).
Press the MENU button. The meter shows INPT .
1.
Press the ∂TARE button. The meter flashes one of the following:
2.
100M (for 0-100 mV dc) (Default)
±50M (for ±50 mV dc)
10V (for 0-10 Vdc)
±5V (for ±5 Vdc)
0-20 (for 0-20 mA dc)
Press the ßNT/GRS button to scroll through available choices.
3.
4.
Press the MENU button to store your choice. The meter momentarily shows
STRD , followed by DEC.P (Decimal Point).