Omega DP25B-S User Manual page 50

Strain gage panel meter
Hide thumbs Also See for DP25B-S:
Table of Contents

Advertisement

4
Configuring The Meter
4.11 USING OUTPUT SCALE AND OFFSET
Press the ∂TARE button. The meter shows last previously stored 4-digit number
11.
(-1999 through 9999) displays with flashing 4th digit.
Press the ßNT/GRS button to change the value of the flashing digit. If you continue
12.
to press the ßNT/GRS button, the flashing digit's value continues to change.
Press the ∂TARE button to scroll to the next digit.
13.
Press the MENU button to store your selection. The meter shows OUT.2 (Output 2).
14.
Note
Press the ∂TARE button. The meter shows selected output.
15.
Note
Press the ßNT/GRS button to change the value of the flashing digit. If you
16.
continue to press the ßNT/GRS button, the flashing digit's value continues to
change.
Press the ∂TARE button to scroll to the next digit.
17.
18.
Press the MENU button to store your selection. The meter momentarily shows
STRD , followed by LK.CF (Lockout Configuration).
40
40
40
This analog signal should correspond to your Read 2 display.
If you select O.2=V for voltage, the maximum signal you may select is 10.00
for an 0-10 Vdc signal output. If you select O.2=C for current, the maximum
signal you may select is 20.00 for a 0-20 or 4-20 mA DC signal output.
WARNING: If the meter displays all flashing values on any item, the value
has overflowed. Press the ßNT/GRS button to start new values.
OT.S.O
(Continued)