8

Specifications

SIGNAL INPUT
Input Ranges:
Isolation:
Noise Rejection:
Resistance:
"Big" Display:
Symbol:
Standard Display:
Symbol:
ANALOG TO DIGITAL
Technique:
Internal resolution:
Read Rate:
ACCURACY AT 25°C
Max Error Strain:
Span Tempco:
Step Response:
Warm Up to Rated Accuracy: 30 min
Excitation Voltage:
Load Regulation:
Line Regulation:
50
SECTION 8. SPECIFICATIONS
0-100 mV, ± 50 mV, 0-10 V, ± 5 V, 0-20 mA, 4-20 mA
Dielectric strength to 2500V transient per 3mm spacing
based on EN 61010 for 260Vrms or DC working voltage
Normal Mode Rejection (NMR) = 60 dB
Common Mode Rejection (CMR) = 120 dB
100 Meg ohms for 100 mV or ±50 mV input range
1 Meg ohm for 10 V or +5 V input range
5 ohms for 20 mA current input range
4-digit, three color programmable 9-segment, LED 21 mm (0.83")
( -1.9.9.9 ~ 9.9.9.9 )
4-digit, 14-segment LED, 13.8 mm (0.54")
( -1.9.9.9. ~ 9.9.9.9. )
Dual slope
15 bits
3/sec Polarity Automatic
±0.03% of reading, ±1 count
50 ppm/°C

1 sec

AC power units
24 V @ 25 mA,
12 V @ 50 mA,
10 V @ 120 mA,
5 V @ 60 mA
1.1%
0.02% per Vac
DC power units
Refer to Table 8-1
Color chart for DC Output
Excitation