Setting The Setpoint 2 Deadband - Omega DP25B-S User Manual

Strain gage panel meter
Hide thumbs Also See for DP25B-S:
Table of Contents

Advertisement

4.9 SETTING THE SETPOINT 2 DEADBAND S2.DB
Note
Refer to Table 6-1 for a summary list of menu configuration.
Setpoint 2 Deadband S2.DB is not active unless your meter has dual relay output
capabilities. The LED's will display whether the S2.DB is active or not. The Setpoint 2
default deadband is 0003. To change the deadband (hysteresis) of Setpoint 2, follow
these steps:
Press the MENU button until the meter shows S2.DB .
1.
Press the ∂TARE button. The meter shows the last previously stored 4-digit
2.
number (0000 through 9999) with flashing 4th digit.
Press the ßNT/GRS button to change the value of the flashing digit. If you
3.
continue to press the ßNT/GRS button, the flashing digit's value continues to
change.
Press the ∂TARE button to scroll to the next digit.
4.
5.
Press the MENU button to store your selection. The meter momentarily shows
STRD , followed by OT.CF (Output Configuration) if you have analog output
capabilities.