Electrical Requirements - AB Sciex 4000 Series Site Planning Manual

Table of Contents

Advertisement

Electrical Requirements

Refer to
Electrical Requirements on page
Requirement
Installation of electrical supplies and fixtures complies with local regulations and safety
standards.
One branch circuit is provided for the mass spectrometer. The mains supply outlet is less
than 1.6 m (63 inches) from the mass spectrometer. One outlet is required. The outlet is
accessible so that the mass spectrometer can be disconnected in an emergency.
One branch circuit is provided for the roughing pump. The mains supply outlet for the
roughing pump is less than 1.6 m (63 inches) from the roughing pump. One outlet is required.
(Optional) One branch circuit with one or more outlets is provided for a gas generator, if
required. Contact the manufacturer of the gas generator for more information.
One branch circuit is provided for the acquisition computer, monitor, and options
the mass spectrometer bench. At least two outlets are required, for the acquisition computer
and monitor.
For LC equipment and other options, contact the manufacturer for more information.
The AC mains supply voltage source does not fluctuate more than ±5% from the nominal
voltage.
Note: Peripheral devices might have different mains supply fluctuation limits. Confirm
the mains supply fluctuation limit with the vendor of each peripheral device that will be
used with the mass spectrometer.
If the voltage is not in the recommended range, a Line Adjustment Transformer (PN WC04179)
is available.
The mains supply includes a correctly installed protective earth conductor.
(Optional) A customer-supplied UPS or power conditioner is provided for the system. The
UPS or power conditioner must deliver 207 VAC to 242 VAC, 50 Hz or 60 Hz, 3200 VA
(minimum). Refer to
Uninterruptible Power Supply or Power Conditioner on
page
20.
A qualified electrician has determined the appropriate AC mains supply configuration based
on the system electrical specifications found in
page
21.
Site Planning Guide
RUO-IDV-01-0563-A | D5040008 A
17.
System Electrical Specifications on
AB SCIEX 4000 Series of Instruments
Site Planning Checklist
Complete
N/A
such as
9 of 33

Hide quick links:

Advertisement

Table of Contents
loading