Ventilation and Waste Collection Requirements

Return to
checklist.
WARNING! Radiation Hazard, Biohazard, or Toxic Chemical Hazard: Make sure
that the mass spectrometer is properly vented and that good general laboratory
ventilation is provided. Adequate laboratory ventilation is required to control
solvent and sample emissions and to provide for safe operation of the mass
spectrometer.
WARNING! Radiation Hazard, Biohazard, or Toxic Chemical Hazard: Use only
qualified personnel for the installation of plumbing and ventilation fixtures,
and make sure that all installations follow local bylaws and biohazard
regulations.
CAUTION: Potential System Damage: Do not connect the source exhaust hose to the vent.
The connection must be made by a qualified FSE.
• Provide a ventilation system with a total negative flow rate capacity of 283 L/min (10 cfm) at item 1 in
A-6.
• Provide a smooth fitting for the roughing pump, with an outside diameter (o.d.) of 3.2 cm (1.25 inches). The
vent must be within 1.5 m (60 inches) of the exhaust port on the roughing pump, and at least 1 m (40 inches)
above the floor. Refer to
• Provide a fitting for the source exhaust drain bottle with an o.d. of 1.3 cm (0.5 inch). The vent must be within
1.5 m (60 inches) of the source exhaust drain bottle. Refer to
Site Planning Guide
RUO-IDV-01-0563-A | D5040008 A
Figure
A-6.
Figure
A-6.
AB SCIEX 4000 Series of Instruments
Site Requirements