Site Requirements
Item
Gas
Gas 1/Gas 2
1
TM
Curtain Gas
CAD gas
2
Source exhaust
3
Note: Use UHP nitrogen (99.999%) or an AB SCIEX-recommended gas generator.
Note: Do not use UHP N
which can damage the emitter tip.
Figure A-5 Swagelok Connection and Tubing

Optional Gas Generators

Return to
checklist.
Gas generators are available from AB SCIEX. Refer to the manufacturer s documentation that comes with the
generator for system specifications. Refer to the manufacturer's documentation for information on the number
and type of outlets required.
AB SCIEX 4000 Series of Instruments
24 of 33
Description
Zero grade air or
N
2
supply;
N
only
2
House air, zero
grade air, or N
2
with the NanoSpray
2
Pressure
100 psi (6.89 bar)
to 105 psi
(7.25 bar)
55 psi (3.79 bar) to
60 psi (4.14 bar)
55 psi (3.79 bar) to
60 psi (4.14 bar)
®
ion source, as there is an increased risk of corona discharge,
Flow Rate
Tubing
(max)
22 L/min
1/4 inch (0.64 cm)
5 L/min for API
1/4 inch (0.64 cm)
TM
4000
systems, 6
L/min for 4000
®
QTRAP
systems
6 L/min
1/4 inch (0.64 cm)