Environmental Requirements - AB Sciex 4000 Series Site Planning Manual

Table of Contents

Advertisement

Site Planning Checklist

Environmental Requirements

Refer to
Environmental Requirements on page
Requirement
An ambient temperature of 15°C to 30°C (59°F to 86°F) is maintained.
Over time, the temperature remains within a range of 4°C (7.2°F), with the rate of the
change in temperature not exceeding 2°C (3.6°F) per hour. Ambient temperature fluctuations
exceeding the limits might result in mass shifts in spectra.
Note: Peripheral devices might have different operating environment requirements.
Confirm the operating environment requirements with the vendor of each peripheral device
that will be used with the system.
Relative humidity is 35% to 50%, non-condensing.
Air conditioning provides a minimum of 8700 Btu/hr (mass spectrometer and roughing pump
only).
BioSafety Requirements
Refer to
BioSafety Requirements on page
Requirement
The site is not designated as BioSafety Level 3 (BSL-3) or BioSafety Level 4 (BSL-4).
AB SCIEX 4000 Series of Instruments
12 of 33