Gas Supply Requirements - AB Sciex 4000 Series Site Planning Manual

Table of Contents

Advertisement

Site Planning Checklist
Electrical Requirements (North America)
Refer to
Electrical Requirements on page
Requirement
Branch circuits for the mass spectrometer and roughing pump are 15 A, 207 VAC to 240
VAC (220 VAC typical), 50 Hz or 60 Hz. Receptacles for these branch circuits are equipped
with CSA/NEMA 6-15R straight-blade receptacles.
The branch circuit for the computer, monitor, and options is 15 A 100 VAC to 240 VAC (120
VAC typical), 50 Hz or 60 Hz.
For requirements for LC equipment, contact the manufacturer.

Gas Supply Requirements

Refer to
Gas Supply Requirements on page
Requirement
Installation of gas supplies and connections complies with local regulations and safety
standards.
Gas 1/Gas 2 is available:
• Zero-grade air or an AB SCIEX-recommended gas generator
• Delivery pressure of 100 psi (6.89 bar) to 105 (7.25 bar) psi with flows up to 22 L/min
TM
Curtain Gas
supply and CAD gas supply is available:
• UHP nitrogen (99.999%) or an AB SCIEX-recommended gas generator
• Delivery pressure of 55 psi (3.79 bar) to 60 psi (4.14 bar) with flows up to 5 L/min for
TM
API 4000
systems or 6 L/min for 4000 QTRAP
Source exhaust gas is available:
• Clean, dry, and oil-free air or UHP nitrogen (99.999%) or an AB SCIEX-recommended
gas generator
• Delivery pressure of 55 psi (3.79 bar) to 60 psi (4.14 bar) with flows up to 6 L/min
(Optional) If an AB SCIEX-recommended gas generator or mass spectrometer bench is used,
then a supply of compressed oil-free air is provided, with a delivery pressure of 120 psi (8.3
bar).
(Optional) If a gas generator without a compressor is used, then a supply of compressed
oil-free air is provided. Contact the manufacturer for more information.
AB SCIEX 4000 Series of Instruments
10 of 33
17.
23.
®
systems
Complete
Complete
Site Planning Guide
RUO-IDV-01-0563-A | D5040008 A