AB Sciex 4000 Series Site Planning Manual page 22

Table of Contents

Advertisement

Site Requirements
Table A-7 Roughing Pump Electrical Specifications: Wet Pump (continued)
Specification
Maximum input current
Maximum input power
Table A-8 Rouging Pump Electrical Specifications: Dry Pump
Specification
Nominal input voltage
Frequency
Maximum input current
Maximum input power
Table A-9 Acquisition Computer Electrical Specifications
Specification
Computer (Dell OptiPlex 9010)
Nominal input voltage
Frequency
Maximum input current
Maximum input power
Monitor (Dell UltraSharp P2314H 23" Widescreen Flat Panel)
Nominal input voltage
Frequency
Maximum input current
AB SCIEX 4000 Series of Instruments
22 of 33
Value
12.5 A
1200 VA
Value
100 VAC to 230 VAC
50 Hz or 60 Hz
13 A (UK), 16 A (Europe), 20 A (US and Japan)
520 VA
Value
100 VAC to 240 VAC
50 Hz or 60 Hz
5A
382 VA
100 VAC to 240 VAC
50 Hz or 60 Hz
2.5 A
Site Planning Guide
RUO-IDV-01-0563-A | D5040008 A

Hide quick links:

Advertisement

Table of Contents
loading